Osteosynthesis of the Displaced Fractures of the Distal Radius Treated with New-Clip® Plate: Case Study in the Basse Terre Hospital

Abstract

Background and Objectives: The distal radius fracture (DRF) is a major public health problem in northern countries. Its frequency is constantly increasing. The locked anterior plate with its well-established biomechanical properties, offers a reliable alternative. The aim of this study was to evaluate the radiological, the functional results and to determine the factors of poor postoperative prognosis of DRF treated with Newclip radial plates®. Methodology: This prospective cohort study evaluates the radiological and functional outcames of displaced radius fractures (DRFs) in patients ≥50 years old treated with Newclip® (locked anterior plates) at the Basse-Terre Hospital in Guadeloupe from 2022 to 2024. The patients were categorized into those with epiphyseal involvement (E1 - E4) and without epiphyseal involvement (E0) based on Laulan’s MEU classification. Radiological parameters (distal radio-ulnar index (DRUI), radial inclination frant view (IRF), radial inclination sagittal view (IRS) were assessed pre and post-operatively. Functional recovery was evaluated at 12 months using the QuickDash questionnaire. Results: Falls were the most common cause of fracture. Post-operatively, SRI was the least restored parameter. Poor prognostic factor for SRI improvement included posterior commimution and unstable fractures. Factors associated with higher QuickDash scores included unstable factures, unrestored DRUI, low plate position, metaphyseal features, and ulnar features. Conclusion: The anterior locking plate osteosynthesis is reliable treatment option with excellent functional outcomes.

Share and Cite:

Cizemba, J. , Albini, H. and Dumontier, C. (2025) Osteosynthesis of the Displaced Fractures of the Distal Radius Treated with New-Clip® Plate: Case Study in the Basse Terre Hospital. Open Journal of Orthopedics, 15, 88-101. doi: 10.4236/ojo.2025.152010.

1. Introduction

Distal radius fractures (DRFs) are the most common fracture injuries, along with proximal femur fractures (which they precede by about ten years) and metacarpal fractures [1]. They are the most associated with osteoporosis, along with those of the spine and proximal femur [2].

Fowler reports that DRFs represent 10% to 25% of all fractures [3]. Cummings et al. in their work on the risks of fractures and coronary heart disease in women note that a 50-year-old white woman has a 16% risk of suffering a hip fracture, a 15% risk of suffering a DRF, and a 32% risk of suffering a vertebral fracture during her remaining life [4].

Figl et al. project an inflation of more than 50% of FRD by 2030 due to population ageing, making FRD a major public health problem [5]. A Congolese DR study ranks this fracture in second position after that of the humeral diaphysis [6].

The DRFs include several types of lesions with different therapeutic modalities and late functional repercussions. Improved knowledge of these lesions is leading to treatment moving towards the search for a reduction that is as anatomical as possible, contained by means that provide sufficient stability to allow early mobilization of the wrist [7] [8].

Several authors have noted an increase in specific surgical treatment adapted to the type of fracture: pinning, external fixator, dorsal plate or anterior plate supplanting traditional orthopedic treatment [2] [3]. Indeed, 45% of fractures of the distal end of the radius are unstable, and require surgical management, the modalities of which remain, however, without consensus to date. The locked anterior plate, which has well-established biomechanical properties, is currently enjoying some popularity and would be a reliable alternative [9] [10].

The aim of this study was to evaluate the subjective functional and immediate postoperative radiological results, and to determine the poor prognostic factors in a prospective series of patients with displaced distal radius fractures treated with the Newclip® locked anterior plate.

2. Material and Methods

Nature, context and period of the study

This prospective cohort study evaluates the radiological and functional outcames of displaced radius fractures (DRFs) in patients ≥50 years old treated with Newclip® (locked antererior plates at the Basse-Terre Hospital in Guadeloupe from 2022 to 2024.

Study population and selection criteria

The patients was categorized into those with epiphyseal involvement (E1-E4) and without epiphyseal involvement (E0) based on Laulan’s MEU classification. Radiological parameters (distal radio-ulnar index (DRUI), radial inclination frant view (IRF), radial inclination sagittal view (IRS) were assessed pre and post-operatively. Functional recovery was evaluated at 12 months using the QuickDash questionnaire.

Inclusion criteria

Patient inclusion took place from 2022 to 2024. Inclusion criteria were all acute fractures (fracture less than 4 weeks old) displaced distal radius, requiring surgical reduction and stabilization with a Newclip® locking plate. Patients should be 50 years of age or older and have been informed in advance of the use of the data for scientific purposes. Functional assessment should be conducted at 12 months by telephone, a return of the questionnaire by email was also an option. Verbal consent was obtained each time. Surgical indications were made by 4 hospital practitioners and they were the main operators in most cases.

Exclusion criteria

Exclusion criteria were difficult-to-use radiological images, fractures and patients who did not complete the QuickDash questionnaire following death or non-response by telephone.

Study parameters

The study parameters were sociodemographic (age, sex, place of residence and circumstance of occurrence); clinical (existence of a diagnosis of FRD; the affected side (right and/or left); preoperative neurological status); Lafontaine instability criteria; radiological (preoperative and postoperative data). Preoperative radiological data were evaluated on computer by an independent evaluator. These data were made with the DRUI, the IRF, the IRS, the existence or not of a posterior comminution; the degree of metaphyseal involvement, the degree of epiphyseal involvement and the possible involvement of the ulna. Carpal involvement was not evaluated. In the postoperative period, all immediate and early complications had to be noted as well as the attitude taken following this complication. Postoperative radiological data were also evaluated on computer by the same evaluator.

Surgical procedure

The surgical procedure was performed under locoregional anesthesia (LRA) alone or general anesthesia (GA) at the request of some patients. In other cases, LRA was supplemented by GA. The operation was performed under a pneumatic tourniquet at the root of the limb, an anterior Henry approach was performed. The plate had to respect the watershed line each time. The reduction of the fracture site was performed by external maneuvers before the plate was placed. The plate was fixed under scopic control (face and profile) to ensure good reduction. The wound was sutured in two planes (subcutaneous and skin), the pronator square was not brought together. The suction drain was optional.

Post-operative protocol

Additional immobilization with a plaster splint was mandatory for a period of 3 weeks. Discharge was authorized as soon as the anesthesia was lifted. Each patient was discharged with analgesic treatment adapted to levels 1 and/or 2. The dressings were done by a state-certified nurse at home. A consultation appointment was scheduled at 3 weeks. Physiotherapy was authorized as soon as the splint was removed. It consisted of gentle passive and active mobilization, physiotherapy for analgesic purposes associated with massage and drainage.

Data source and measurements

Data was collected using a pre-established form containing the study parameters. Data was entered into Excel 2013 and then transferred to SPSS (Statistical package for the social sciences) version 27 and R 5.5. The radiological analysis was performed directly on digitised images using PACS (Picture Archiving and Communication Systems) software. A non-probability sampling was used.

Statistical analyses

It was carried out with the SPSS software version 27 and R 5.5. To compare the quantitative and qualitative values, we used as tests of statistical analyses:

  • The Student t test, for the comparison of quantitative variable means.

  • The chi-2 test comparison of qualitative variables.

  • The ANCOVA analysis of variance test, for the comparison of inter- and intragroup effects.

  • Finally, the analysis of regressions and correlations.

  • The significance level was set at p < 0.05.

Ethical consideration

This study had received permission from the Head of Department. We did not need authorization from the Committee for the Protection of Persons because the research was non-interventional.

The data were processed anonymously in accordance with the Declaration of Helsinki.

3. Results

A total of 61 patients was initially included, 8patients (13% of the initial sample) were excluded. We retained 53 patients. The majority were women: 33 women (62.3%). Among our patients, 10 (18.9%) were tourists. The 62 - 67 age group predominated, 30.2% (n = 16). Falls were the most frequent circumstance of occurrence in our study, n = 48 (90.5) (Table 1). Our results showed that the left side was the most affected n = 31(58.5%). Fractures were unstable in 66% n = 35. The distribution of patients according to epiphyseal damage showed that 28 (52.8%) had no epiphyseal damage, compared with 25 (47.2%).

Table 1. Evaluation of preoperative radiological indices.

Variables

Disturbed

(frequency N = 53)

Mild

Moderate

Severe

Exaggerated

Median (extreme)

Preoperative IRUD

41 (77.5%)

27 (50.9%)

-

14 (26.6%)

-

2.4 (−2 and 8)

Preoperative IRF

41 (77.4%)

16 (39%)

13 (31.7%)

8 (19.5%)

4 (9.7%)

16 (5 and 36)

Preoperative IRS

40 (75.5%)

7 (17.5%)

3 (7.5%)

9 (22.5%)

21 (52.5%)

16 (−35 and 4)

The preoperative assessment of the DRUI (Table 2) revealed that: the DRUI was disturbed in 41 patients (77.5%), with a median of 2.4 (−2 to 8). The FRI was also disturbed in 77.5% n = 41 with a median of 16 (5 - 36). The IRS was disturbed in 75% n = 40 with a median of 16 (−35 to 44). The DRUI was severely impaired in the E0 group n = 23 and this disruption was often mild (˂3 mm) in both E0 groups 69.6% (n = 16) and 61.1% (n = 11) for the E1-E4 group; p = 0.571.

The FRI was disturbed in 23 patients in the E0 group and this disturbance was largely mild (≤5˚) p = 0.571. The IRS was disturbed in 22 patients in the E0 group, and this disturbance was largely exaggerated (˃25˚): E0 14 (63.6); E1 - E4 7 (39.9) p = 0.933.

Table 2. Evaluation of postoperative radiological indices and distribution of the degree of loss in the two groups.

1. DRUI

postoperative

Frequency

%

not recovered

6

11.30%

Degree of lossn = 6

E0

E1-E4

Total

Slight

3 (50%)

3 (50%)

6

34

64.20%

Severe

0 (0%)

0 (0%)

0

Recovered

13

24.50%

normal at beginning

53

100%

Total

2. IRF

E0

E1-E4

Degree of lossn = 12

E0

E1-E4

Total

P

not recovered

8 (28.6%)

4 (16%)

Slight

6 (50%)

3 (50%)

9

0.002

moderate

1 (8.3%)

1 (8.3%)

2

Severe

1 (8.3%)

0

1

15(53.6%)

14 (56%)

Total

8 (66.7%)

4 (33.3%)

12 (100%)

Recovered

5 (17.9%)

7 (28%)

normal at beginning

28 (100%)

25 (100%)

Total

3. IRS

E0

E1-E4

Degree of lossn = 30

E0

E1-E4

Total

p

not recovered

15 (53.5%)

15 (60%)

Slight

5 (16.6%)

5 (16.6%)

10 (30.3%)

0.007

moderate

4 (13.3%)

2 (6.6%)

6

Severe

6 (20%)

8 (26.7%)

14

8 (28.6%)

7 (28%)

Total

15 (49.9%)

15 (49.9%)

30 (100%)

Recovered

5 (17.9%)

3 (12%)

normal at beginning

Post-operative results (Table 3) showed an absence of DRUI correction in 11.3% (n = 6). This lack of correction was slight (˂3 mm) in both groups, with equal proportions. Only 11 patients, or 21%, had their FRI uncorrected. The loss was milder (˂5˚) in 17% (p = 0.002).

Table 3. Analysis of Quick Dash score variant.

Group of patients

N

median

extreme

Lower limit

Higher limit

minima

maxima

E0

28

13.56

0-57

8.4

18.73

0

57

E1-E4

25

221.96

0-80

14.26

29.55

0

80

Total

53

17.5

0-80

12.99

22.01

0

80

Varience homogeneity test

Quick dash 1 year

Levenede stat

dd1

dd12

Sig

based on average

15 (53.6%)

Based on median

5 (17.9%)

1

51

0.012

Based on median with adjusted ddl

0.036

1

37.677

0.851

based on troncated mean

0.094

1

51

0.76

ANCOVA

Sum of squares

dd1

mean square

F

Sig

Intergroups

919.477

1

919.477

3.602

0.03

intragroups

13017.643

51

255.248

Total

13937.12

52

The IRS was not restored in 53.5% (n = 30), in almost equal proportions in both groups.

In terms of poor prognostic factors, patients over 50 years of age were 1.003 times more likely to experience no postoperative improvement in FRI. Patients with epiphyseal features were 6.204 times more likely to have an unimproved FRI. Patients with unstable fractures were 1.518 times more likely to have unimproved FRI. The variables selected explained 92% of the phenomenon and were globally significant (Table 4).

Table 4. Explanation for the lack of improvement in IRF.

Non Coef.

Amelioration of IRF

St. Err

t-value

p value

95% Conf

Interval

Sig

Age

1.003

0.037

0.09

0.03

0.934

1.078

**

patient’s sex

female

0.953

0.728

−0.06

0.005

0.213

4.262

***

patient group

E1-E4

6.204

5.452

2.08

0.038

1.108

34/733

**

posterior communition

yes

5.085

6.552

1.26

0.007

0.407

63.532

***

Lafontaine criteria

Unstable fracture

1.518

1.94

0.33

0.044

0.124

18.571

**

Constant

0.129

0.317

−0.71

0.076

0

35.725

*

Mean dependent var

0.421

0.5

Pseudo r-squared

0.92

Number of

obs

38

Chi-square

7.36

Prob > chi2

0/095

Akaike crit. (AIC)

56.368

Bayesan

crit (BIC)

66.193

***p < 0.01, **p < 0.05, *p < 0.1.

Regarding the IRS, patients with posterior comminution were 3.16 more likely to experience no improvement in IRS. Those with unstable fractures were 2.453 more likely to get IRS. The selected variables explained 81% of the phenomenon and were globally significant (Table 5).

Table 5. Explanation for the lack of improvement in IRS.

No Amelioration of IRS

Coef.

St. Err

t-value

P value

95% Conf

Interval

Sig

Age

1.323

0.084

2.01

0

0.121

1.164

***

patient’s sex:

female

2.607

1.308

2.84

0.004

0.421

5.004

***

patient group

E1-E2-E3-E4

2.508

2.054

4.04

0.069

659

4.809

*

posterior communition

yes

3.135

3.092

3.07

0.035

0.201

3.879

**

Lafontaine criteria

Unstable fracture

2.453

2.001

2.01

0.027

0.101

8.23

**

Constant .508

0.493

7.05

0.001

4.118

***

Mean dependent var

0.868

SD dependent

var

1.303

Pseudo r-squared

0.81

number of

obs

38

Chi-square

2.604

Prob > chi2

0.03

Akaike crit. (AIC)

8.999

Bayesian cri

t. (BIC)

2.705

***p < 0.01, **p < 0.05, *p < 0.1.

Explanatory analysis of the QuickDash score

The median Quick Dash in the E0 group was low: 13.56 (0 - 57) compared with the Quick Dash in the E1 - E4 group: 21.91 (0 - 80). In bivariate analysis, we noted an association between Quick Dash, epiphyseal feature, ulnar feature and IRS p < 0.05%. In multivariate analysis, it is apparent that the unstable fracture, the unrestored postoperative DRUI, the low position of the plate in relation to the pronator squared line, the M1-M4 metaphyseal features, the existence of C1 - C4 ulnar features contributed positively to the Quick Dash score, and this contribution was significant at the 5% threshold (Table 6).

Table 6. Quickdash poor prognosis factors.

Score dash

Coef.

St. Err

t-value

p value

95% Conf

Interval

Sig

Patient’s sex

Masculine

−793

4.788

−0.17

0.069

10,463

8878

*

State of DRUI

0

post op

.

not healed

8.559

7.312

−1.17

49

3326

16209

**

Position of the plate under

3262

4.916

−0.66

0.011

2.19

4.665

**

State of IRF post-op

not healed

6157

5.367

1.15

8

0.4681

6995

***

State of IRS post-op

not healed

7976

4.59

1.74

0.09

0.1293

7246

*

Ulnar feature: C1 - C4

6683

4.739

−1.41

0.006

5.253

12888

***

Metaphysal feature

M3

10.289

6.833

1.51

4

−3.511

24.09

**

M4

8702

5217

1.67

0.003

−1834

19 237

***

Lafontaine criteria

Unstable fracture Group

1213

5.09

0.24

13

−65

1492

**

E-E4

7728

4633

1.67

3

−1629

17084

***

Constant

11257

9726

1.16

254

−8386

30.9

Mean dependent var

17702

SD dependent

var

16452

R-squared

0.773

number of

obs

53

F-test

7.541

Prob > F

0.002

4. Discussion

Strength of the study

This is a prospective study conducted in a university-oriented center.

The objectives determined for this study were achieved.

The number of patients included in this series (53 patients) is close to the large series encountered in the literature. The number of patients lost to follow-up was not significant.

Our series has integrated, in addition to the Laulan classification, the elements of Lafontaine instabilities.

It constitutes the first large cohort study carried out in the orthopedics department of the CHBT.

All patients benefited from the same plate placed using the same technique, limiting the biases caused by the installation of different materials.

They were all reassessed by the same assessor at 12 months with a low complication rate.

Our study has certain limitations:

  • The variability of radiological measurements. These measurements are also very dependent on the quality of the X-ray images, which itself depends on the medical electroradiology manipulators and the patient’s positioning possibilities. Nevertheless, precise measurements were made, leaving no room for a simple visual estimation.

  • Another limitation lies in the fact that the surgical procedure was performed by a team of 4 practitioners with different experiences, which can modify the results.

  • This series did not evaluate wrist movements or grip strength.

  • The study also did not assess patient satisfaction

Our cohort

As previously stated, DRFs are the most common fracture lesions along with proximal femur fractures (which they precede by about ten years) [1]. Our sample revealed an annual frequency of 27 anterior locked plate osteosyntheses. The data in the literature in our possession note a significant use of this procedure in the treatment of FRD. Huttinen [2] found that between 1998 and 2008, the use of surgical treatment in the treatment of FRD had doubled the reasons for this trend have remained unclear. They also found an increase in the use of internal fixation compared to other techniques, which more than doubled over an 11-year period.

Bjarke et al., without determining the exact annual frequency of the use of plates in the management of DRFs, noted in their study that the use of plates had become the first choice since 2007 with good clinical and radiological results. The same authors noted that its use reached a rate of 9% in 2018 [11].

Koval, for his part, noted in his study a decrease in broaching in favor of plates due to the high rate of complications following broaching [12].

Our study population reveals that the age group of 62 - 67 years was the most preponderant with 30.2% (n = 16). The mean age is 64.89 ± 10.21 years (p = 0.004). This mean is close to that found by Jakub (64 ± 17.9 years) in his study devoted to the quality of life after a FRD [13]. It is slightly lower than that found by Mayank et al (71.1 ± 8.89) [14]. and to that found by Maxence et al. (72 years). The average age of the women was higher than that of the men: 67.35 ± 11.9. This is in line with several series published in the literature (Chen et al 2015, Stenley 2016) [10] [15]. When both groups were considered, the epiphyseal feature was observed in patients aged over 60 years 44.2% (n = 23) and female.

In terms of distribution, DRFs are distributed as a function of age, which is bimodal. There is a first peak due to high-energy trauma in young individuals (5 - 24 years) with good bone quality. The second peak occurs mainly after the age of 50. It follows low-energy trauma in a population composed mainly of post-menopausal women with osteoporotic bone [10] [15]. The bone density in the latter population is often reduced. The association between DRFs and bone mineral density in postmenopausal women is well established [16].

Our results also showed that falls were the most frequent circumstance of occurrence n = 48 (90.5%). The falls found in our series meet the series of several authors [17]-[19].

With regard to radiological parameters, our series evaluated 3 radiological parameters. Herzberg and Dumontier have provided a simple way to understand the DRFs, allowing to identify and therefore treat all the injuries without forgetting any. This is a list of essential elements whose anatomy must be restored, because it is linked to the functional prognosis. These are the IRS, the FRI, the joint impaction (IA) and the shortening of the radius (by metaphyseal compression) as well as the intracarpal and distal radio-ulnar joint (DRUJ) lesions contemporary with the trauma [20].

Our preoperative results showed a disturbance of the IRS in 76% of cases. Postoperatively, the IRS was the least restored index in half the cases. This lack of reduction was 60% in the E1-E4 group and 53% in the E0 group (p = 0.819). This poor post-operative result corresponds to that of the series by Mignemi et al, in a retrospective series of 185 patients, these authors found disappointing radiological results after DRFs osteosynthesis. In fact, a normal IRS (11 degrees) was restored in only 48% of cases [21]. Jeudy et al., in evaluating their prospective series of 41 DRFs treated with locking anterior plates, noted losses of angular reduction for glenoid anteversion occurring mainly in the first post-operative weeks. Most of them were minimal, of the order of a few degrees, and mainly concerned posterior displacement fractures [22].

In our series, the loss of inclination was slight at 16.9% (n = 5), moderate at 14.3% (n = 4) and severe at 21.4% (n = 6) for the E0 group. For the E1-E4 group, slight loss accounted for 20% (n = 5), moderate for 8% (n = 2) and severe for 32% (n = 8) (p = 0.758). Fowler et al, on the other hand, in a prospective series published in 2013 on 37 patients with an average age of 57 years, found at the end of 14 months of follow-up no statistically significant difference with regard to IRS [3]. Our work also highlighted two poor prognostic factors: posterior comminution and the existence of at least three of Lafontaine’s criteria. We believe that posterior communication remains an important element to consider when restoring IRS. Other authors propose the placement of a graft [23].

The absence of IRS restoration is described as the parameter with the greatest functional impact in particular the increase in pressures at the level of the dorsal surface of the radioulnar joint and the styloid, increase in pressures on the scaphoid as well as midcarpal instability as soon as the radial glenoid is horizontalized [24]-[26].

Our study found an association between Quick Dash score and absence of SRI restoration, and the difference was statistically significant p = 0.027. Karnezis in his work did not find a significant correlation between loss of SRI and the PRWE subscore [27].

Preoperatively, 82.1% of patients in the E0 group and 77.4% in the E1-E4 group had a disturbed FRI. Postoperatively, only 11 patients (21%) had their FRI uncorrected. The results of our series are close to those of Jeudy et al. [22], and better than those of Mignemi et al. [21], whose results were very disappointing: normal FRI (22 degrees) was restored in only 43% of patients. Jeudy et al. [22] report that most losses in their series were minimal. In our series, loss was slight in 17%, moderate in 3.8% and severe in 1%.

Few clinical studies have focused on FRI. Jenkins showed that loss of 7˚ resulted in a decrease in flexion [28]. In our series, we found no correlation between Quick Dash score and absence of FRI/H recovery p = 0.169. This observation is in line with the study by Garland et al, who demonstrated that an abnormal radial angle does not affect the final functional outcome after DRF [25]. Karnezis [27] found no correlation between radial angle loss and PREW score. However, in the literature we have, we did not find any factors that could explain postoperative RFI loss. In our study, age over 50, the existence of an epiphyseal line and unstable fractures were identified as unfavourable factors.

Concerning DRUI, the preoperative result revealed that it was disturbed in 82.1% (n = 23) of the E0 group and in 72% (n = 18) of the E1-E2 group. The post-operative result was marked by an absence of correction in 11.3% of cases, n = 6, and this loss was less than 3 mm in all cases. This result is better than that of Jeudy et al. who, in their prospective series of 41 DRFs treated with locking anterior plate, noted loss of reduction, with a change in DRUI in 29% of cases [22]. The result of Mignemi et al. [21] was still very poor, with loss of DRUI observed in 47%.

Normal DRUI is either negative or neutral, while a positive index indicates radial shortening. Radial shortening, secondary to metaphyseal comminution, is the second most important arthrogenic factor after joint impingement, it can cause several repercussions, in particular an increase in pressure on the scaphoid fossa of the radius with an impact on the distal radioulnar joint, importance of the change in pressure distribution as soon as the radius presents a shortening of more than 2.5 mm compared to the ulna [29] [30].

McQueen reported that a shortening of the radius of more than 2 mm significantly led to a worsening of symptoms and a decrease in strength [31].

The assessment of the DASH score at 12 months in both patient groups revealed a low median of 13.56 (extreme 0 - 57) for E0 patients and a slightly high median of 21.91 (0 - 89) for E1-E4 patients (p = 0.26).

The results of our series are in line with those of several other studies of anterior locking plates with a 12-month follow-up [27] [28]. Our average is slightly higher than that found by Antonio et al. These authors analysed the DASH score in a group of 501 patients, including 133 surgically treated patients. They found a low Quick Dash average in both groups at one-year follow-up (7.5), and older age was the only factor for the poor outcome [29]. In our study, we found 4 poor prognostic factors: the existence of at least three Lafontaine criteria, the low position of the plate in relation to the pronator quadratus line, the M3 metaphyseal feature, the loss of DRUI.

5. Conclusion

DRFs of the radius are on the increase. Their treatment has undergone major changes over the last 20 years, with a tendency towards surgery. This fracture is closely linked to osteoporosis. Despite the controversy and lack of consensus in the treatment of these fractures, anterior locking plate osteosynthesis has recently gained in popularity. Its extension to fractures previously eligible for pinning has been noted. Anterior locking plate osteosynthesis thus represents a reliable therapeutic alternative with an excellent functional outcome, which appears to be earlier than after pinning. Nevertheless, the risk of loss of reduction has also been observed by other authors at different times. In our study, loss of reduction was observed immediately post-operatively, and poor prognostic factors were identified.

Authors’ Contributions

Cizemba Johnny: substantial contribution to design, configuration, data acquisition. Dumontier Christian and Albini Hugues: substantial contribution to the conception and direction of the work. All the authors have read and approved the final version of the manuscript.

Conflicts of Interest

We declare no conflict of interest.

References

[1] Court-Brown, C.M. and Caesar, B. (2006) Epidemiology of Adult Fractures: A Review. Injury, 37, 691-697.[CrossRef] [PubMed]
[2] Huttinen, T.T. (2014) Surgical Traitement of Upper Limb Fractures An Evidence-Based Approach. University of Tampere.
[3] Fowler, J.R. and Ilyas, A.M. (2013) Prospective Evaluation of Distal Radius Fractures Treated with Variable-Angle Volar Locking Plates. The Journal of Hand Surgery, 38, 2198-2203.[CrossRef] [PubMed]
[4] Cummings, S.R. (1989) Lifetime Risks of Hip, Colles’, or Vertebral Fracture and Coronary Heart Disease among White Postmenopausal Women. Archives of Internal Medicine, 149, 2445-2448.[CrossRef] [PubMed]
[5] Figl, M., Weninger, P., Liska, M., Hofbauer, M. and Leixnering, M. (2009) Volar Fixed-Angle Plate Osteosynthesis of Unstable Distal Radius Fractures: 12 Months Results. Archives of Orthopaedic and Trauma Surgery, 129, 661-669.[CrossRef] [PubMed]
[6] Cizemba, J., Bakumobatane, L.M., Mutombo, E.K., Mukendi, M.K., Kabengele, R.M., Bamuleka, D.M., et al. (2023) Upper Limb Fractures in Kinshasa: Therapeutic Approach in a Sub-Saharan African Country. Open Journal of Orthopedics, 13, 405-420.[CrossRef]
[7] Bradway, J.K., Amadio, P.C. and Cooney, W.P. (1989) Open Reduction and Internal Fixation of Displaced, Comminuted Intra-Articular Fractures of the Distal End of the Radius. The Journal of Bone & Joint Surgery, 71, 839-847.[CrossRef]
[8] Boretto, J.G., Gallucci, G.L., Alfie, V.A., Donndorff, A. and De Carli, P. (2009) Les fractures intraarticulaires du radius distal à déplacement postérieur traitées par plaque palmaire à stabilité angulaire. Technique et résultats. Chirurgie de la Main, 28, 18-25.[CrossRef] [PubMed]
[9] Egol, K.A., Kubiak, E.N., Fulkerson, E., Kummer, F.J. and Koval, K.J. (2004) Biomechanics of Locked Plates and Screws. Journal of Orthopaedic Trauma, 18, 488-493.[CrossRef] [PubMed]
[10] Stenley, L. (2016) Fractures de l’extrémité distale du radius ostéosynthésées par plaque antérieure verrouillée: Étude prospective de 81 patients. Master’s Thesis, Université de Picardie Jules Verne.
[11] Viberg, B., Tofte, S., Rønnegaard, A.B., Jensen, S.S., Karimi, D. and Gundtoft, P.H. (2023) Changes in the Incidence and Treatment of Distal Radius Fractures in Adults—A 22-Year Nationwide Register Study of 276,145 Fractures. Injury, 54, Article ID: 110802.[CrossRef] [PubMed]
[12] Koval, K.J., Harrast, J.J., Anglen, J.O. and Weinstein, J.N. (2008) Fractures of the Distal Part of the Radius. The Journal of Bone and Joint Surgery-American Volume, 90, 1855-1861.[CrossRef] [PubMed]
[13] Marchewka, J., Marchewka, W. and Golec, E. (2021) Quality of Life after Distal Radius Fractures. Rehabilitacja Medyczna, 24, 13-18.[CrossRef]
[14] Jayaram, M., Wu, H., Yoon, A.P., Kane, R.L., Wang, L. and Chung, K.C. (2023) Comparison of Distal Radius Fracture Outcomes in Older Adults Stratified by Chronologic vs Physiologic Age Managed with Casting vs Surgery. JAMA Network Open, 6, e2255786.[CrossRef] [PubMed]
[15] Chen, C.H., Zhou, R.K., Zhen, H.Q., Huang, L. and Jiao, Y.J. (2015) Efficacy of Volar and Dorsal Plate Fixation for Unstable Dorsal Distal Radius Fractures. International Journal of Clinical and Experimental Medicine, 8, 4375-4380.
[16] Lafontaine, M., Delince, P., Hardy, D. and Simons, M. (1989) Instability of Fractures of the Lower End of Radius: A Propos of a Series of 167 Cases. Acta Orthopaedica Belgica, 55, 203-216.
[17] Hanane, A. (2008) Les fractures distales du radius, analyse radiologique. Université cadi Ayyad Marrakesh.
[18] Chung, K.C. and Spilson, S.V. (2001) The Frequency and Epidemiology of Hand and Forearm Fractures in the United States. The Journal of Hand Surgery, 26, 908-915.[CrossRef] [PubMed]
[19] Dahmi, Z. (2009) Le traitement chirurgical des fractures de l’extrémité distale du ra-dius. Revue de Chirurgie Orthopédique et Traumatologique, 39, 18-23.
[20] Dumontier, C. and Herzberg, G. (2001) Classifications du radius distal. Revue de Chirurgie Orthopédique et Traumatologique, 87, S78-S85.
[21] Mignemi, M.E., Byram, I.R., Wolfe, C.C., Fan, K., Koehler, E.A., Block, J.J., et al. (2013) Radiographic Outcomes of Volar Locked Plating for Distal Radius Fractures. The Journal of Hand Surgery, 38, 40-48.[CrossRef] [PubMed]
[22] Jeudy, J., Steiger, V., Boyer, P., Cronier, P., Bizot, P. and Massin, P. (2012) Treatment of Complex Fractures of the Distal Radius: A Prospective Randomised Comparison of External Fixation ‘versus’ Locked Volar Plating. Injury, 43, 174-179.[CrossRef] [PubMed]
[23] Obert, L., Rey, P., Uhring, J., Gasse, N., Rochet, S., Lepage, D., et al. (2013) Ostéosynthèse des fractures du radius distal chez l’adulte: Mise au point. Revue de Chirurgie Orthopédique et Traumatologique, 99, 174-191.[CrossRef]
[24] Short, W.H., Palmer, A.K., Werner, F.W. and Murphy, D.J. (1987) A Biomechanical Study of Distal Radial Fractures. The Journal of Hand Surgery, 12, 529-534.[CrossRef] [PubMed]
[25] Taleisnik, J. and Watson, H.K. (1984) Midcarpal Instability Caused by Malunited Fractures of the Distal Radius. The Journal of Hand Surgery, 9, 350-357.[CrossRef] [PubMed]
[26] Kihara, H., Palmer, A.K., Werner, F.W., Short, W.H. and Fortino, M.D. (1996) The Effect of Dorsally Angulated Distal Radius Fractures on Distal Radioulnar Joint Congruency and Forearm Rotation. The Journal of Hand Surgery, 21, 40-47.[CrossRef] [PubMed]
[27] Karnezis, I.A., Panagiotopoulos, E., Tyllianakis, M., Megas, P. and Lambiris, E. (2005) Correlation between Radiological Parameters and Patient-Rated Wrist Dysfunction Following Fractures of the Distal Radius. Injury, 36, 1435-1439.[CrossRef] [PubMed]
[28] Jenkins, N.H. and Mintowt-Czyz, W.J. (1988) Mal-Union and Dysfunction in Colles’ Fracture. Journal of Hand Surgery, 13, 291-293.[CrossRef] [PubMed]
[29] Pogue, D.J., Viegas, S.F., Patterson, R.M., Peterson, P.D., Jenkins, D.K., Sweo, T.D., et al. (1990) Effects of Distal Radius Fracture Malunion on Wrist Joint Mechanics. The Journal of Hand Surgery, 15, 721-727.[CrossRef] [PubMed]
[30] Werner, F.W., Glisson, R.R., Murphy, D.J. and Palmer, A.K. (1986) Force Transmission Trough the Distal Radioulnar Carpal Joint: Effect of Ulnar Lengthening and Shortening. Handchirurgie, 18, 304-308.
[31] McQueen, M. and Caspers, J. (1988) Colles Fracture: Does the Anatomical Result Affect the Final Function? The Journal of Bone and Joint Surgery. British volume, 70, 649-651.[CrossRef] [PubMed]