<?xml version="1.0" encoding="UTF-8"?><!DOCTYPE article  PUBLIC "-//NLM//DTD Journal Publishing DTD v3.0 20080202//EN" "http://dtd.nlm.nih.gov/publishing/3.0/journalpublishing3.dtd"><article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" dtd-version="3.0" xml:lang="en" article-type="research article"><front><journal-meta><journal-id journal-id-type="publisher-id">OJAnes</journal-id><journal-title-group><journal-title>Open Journal of Anesthesiology</journal-title></journal-title-group><issn pub-type="epub">2164-5531</issn><publisher><publisher-name>Scientific Research Publishing</publisher-name></publisher></journal-meta><article-meta><article-id pub-id-type="doi">10.4236/ojanes.2016.67017</article-id><article-id pub-id-type="publisher-id">OJAnes-68780</article-id><article-categories><subj-group subj-group-type="heading"><subject>Articles</subject></subj-group><subj-group subj-group-type="Discipline-v2"><subject>Medicine&amp;Healthcare</subject></subj-group></article-categories><title-group><article-title>
 
 
  Regardless of the Puncture Angle, a Tuohy Needle with Sand-Blasted Steel at the Tip of the Bevel Is a Valid Needle on the Ultrasonic Images
 
</article-title></title-group><contrib-group><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Yuki</surname><given-names>Ichikawa</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Hironobu</surname><given-names>Ueshima</given-names></name><xref ref-type="aff" rid="aff2"><sup>2</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Hiroshi</surname><given-names>Otake</given-names></name><xref ref-type="aff" rid="aff2"><sup>2</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Akira</surname><given-names>Kitamura</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref></contrib></contrib-group><aff id="aff2"><addr-line>Department of Anesthesiology, Showa University Hospital, Tokyo, Japan</addr-line></aff><aff id="aff1"><addr-line>Department of Anesthesiology, Saitama International Medical Center, Saitama, Japan</addr-line></aff><pub-date pub-type="epub"><day>21</day><month>07</month><year>2016</year></pub-date><volume>06</volume><issue>07</issue><fpage>101</fpage><lpage>104</lpage><history><date date-type="received"><day>29</day>	<month>June</month>	<year>2016</year></date><date date-type="rev-recd"><day>accepted</day>	<month>18</month>	<year>July</year>	</date><date date-type="accepted"><day>21</day>	<month>July</month>	<year>2016</year></date></history><permissions><copyright-statement>&#169; Copyright  2014 by authors and Scientific Research Publishing Inc. </copyright-statement><copyright-year>2014</copyright-year><license><license-p>This work is licensed under the Creative Commons Attribution International License (CC BY). http://creativecommons.org/licenses/by/4.0/</license-p></license></permissions><abstract><p>
 
 
  Background: A high ability to visualize the needle on the ultrasonic images is necessary to perform the ultrasound-guided nerve block safely. The Rafa Tuohy needle
  <sup>R</sup> (Vygon, Paris, France, Rafa) is a non-stimulating Tuohy needle with sand-blasted steel at the tip of the bevel. We examined the degree to which the Rafa enhanced the visibility of ultrasonic images compared with the non-coated Tuohy needle. Methods: We punctured the Blue Phantom. The dimensions of both the Rafa and the non-coated Tuohy needles were 18 G &#215; 80 mm. The puncture angle is 30 degrees and 45 degrees from the Blue Phantom. We measured the intensity of the tip of the bevel at a depth of 0.5, 1.0, 1.5 and 2.0 cm for the puncture angle of 30 degrees, and 0.5, 1.0, 1.5, 2.0, 2.5 and 3.0 cm for the puncture angle of 45 degrees. Six anesthesiologists with more than seven years of experience performed three punctures using each needle. Results: As an outcome, we concluded that at a puncture angle of both 30 degrees and 45 degrees, the intensity of the non-coated Tuohy needle decreased with the depth. On the other hand, at an angle of 30 degrees, the intensity of the Rafa needle did not decrease, and at an angle of 45 degrees the intensity only decreased very slightly as the depth increased. Conclusions: The Tuohy needle with sand-blasted steel at the tip of the bevel provided greater visualization than the non-coated Tuohy needle on the ultrasound images.
 
</p></abstract><kwd-group><kwd>Puncture Angle</kwd><kwd> Sand-Blasted Steel</kwd><kwd> Ultrasonic Images</kwd></kwd-group></article-meta></front><body><sec id="s1"><title>1. Introduction</title><p>When we perform the ultrasound-guided nerve blocks, we will encounter complications such as nerve or organ injuries, and local anesthetic systemic toxicity [<xref ref-type="bibr" rid="scirp.68780-ref1">1</xref>] - [<xref ref-type="bibr" rid="scirp.68780-ref3">3</xref>] . One of the reasons, which can cause some complications, is the poor visualization on the ultrasound images in performing the ultrasound-guided nerve blocks. Therefore, the needle has been devised in various ways in order to improve the visualization of the needle on the ultrasound images [<xref ref-type="bibr" rid="scirp.68780-ref4">4</xref>] - [<xref ref-type="bibr" rid="scirp.68780-ref7">7</xref>] . The Rafa Tuohy needle<sup>R</sup> (Vygon, Paris, France, Rafa) is a non-stimulating Tuohy needle with sand-blasted steel at the tip of the bevel. We can predict that the modification would also enhance the visibility of ultrasonic images regardless of some puncture angles. In this study, we examined the degree to which the Rafa enhanced the visibility of ultrasonic images compared with the non-coated Tuohy needle.</p></sec><sec id="s2"><title>2. Materials &amp; Methods</title><p>We did not submit this study to an institutional review board due to lack of human testing with the Blue Phantom. We prepared the S-Nerve ultrasound system (SonoSite FUJIFILM, Washington, America) and the Blue Phantom for the peripheral nerve blocks (CAE Healthcare, Sarasota, FL). We punctured the Blue Phantom by using a 50-x high-frequency linear probe with a needle guide kit (Civco Medical Instrument Co., Orange City, IA). The dimensions of both the Rafa and the non-coated Tuohy needles were 18 G &#215; 80 mm. The puncture angle was 30 degrees and 45 degrees to the skin. We used Image J software, which is an image-processing program designed for scientific multidimensional images, to measure the intensity of the tip of the bevel at a depth of 0.5, 1.0, 1.5 and 2.0 cm for the puncture angle of 30degrees, and 0.5, 1.0, 1.5, 2.0, 2.5 and 3.0 cm for the puncture angle of 45 degrees. Six anesthesiologists with more than seven years of experience performed the ultrasound-guided nerve block and made three punctures using each needle.</p></sec><sec id="s3"><title>3. Statistics</title><p>The Student’s t-test was used for statistical analysis, and significance was accepted at p &lt; 0.05. The numerical variables of the intensity at the tip of the needle are expressed as mean (standard deviation) (cd/m<sup>2</sup>).</p></sec><sec id="s4"><title>4. Results</title><p>We showed the intensity of two Tuohy needles at each point in <xref ref-type="table" rid="table1">Table 1</xref> and <xref ref-type="table" rid="table2">Table 2</xref>. <xref ref-type="table" rid="table1">Table 1</xref> was the intensity of the puncture angle set at 30 degrees and the intensity of Rafa was 194.7 (6.6), 183.5 (5.5), 190.1 (11.3), 182.8</p><table-wrap id="table1" ><label><xref ref-type="table" rid="table1">Table 1</xref></label><caption><title> The intensity of each needles for the puncture angle of 30 degrees (n = 18). Data are expressed as mean (standard deviation) (cd/m<sup>2</sup>). Rafa = Rafa Tuohy needle. Non-coated = Non-coated Tuohy needle. On the all point, the intensity of the Rafa was clearly higher than the intensity of the non-coated needle (P &lt; 0.0001)</title></caption><table><tbody><thead><tr><th align="center" valign="middle"  rowspan="2"  ></th><th align="center" valign="middle"  colspan="4"  >The depth from Blue Phantom (cm)</th></tr></thead><tr><td align="center" valign="middle" >0.5</td><td align="center" valign="middle" >1.0</td><td align="center" valign="middle" >1.5</td><td align="center" valign="middle" >2.0</td></tr><tr><td align="center" valign="middle" >Rafa</td><td align="center" valign="middle" >194.7 (6.6)</td><td align="center" valign="middle" >183.5 (5.5)</td><td align="center" valign="middle" >190.1 (11.3)</td><td align="center" valign="middle" >182.8 (5.5)</td></tr><tr><td align="center" valign="middle" >Non-coated</td><td align="center" valign="middle" >134.8 (4.2)</td><td align="center" valign="middle" >116.8 (7.9)</td><td align="center" valign="middle" >94.5 (4.4)</td><td align="center" valign="middle" >67.6 (1.4)</td></tr></tbody></table></table-wrap><table-wrap id="table2" ><label><xref ref-type="table" rid="table2">Table 2</xref></label><caption><title> The intensity of each needles for the puncture angle of 45˚ (n = 18). Data are expressed as mean (standard deviation) (cd/m<sup>2</sup>). On the all point, the intensity of the Rafa was clearly higher than the intensity of the non-coated needle (P &lt; 0.0001). Rafa = Rafa Touhy needle. Non-coated = Non-coated Touhy needle</title></caption><table><tbody><thead><tr><th align="center" valign="middle"  rowspan="2"  ></th><th align="center" valign="middle"  colspan="6"  >The depth from Blue Phantom (cm)</th></tr></thead><tr><td align="center" valign="middle" >0.5</td><td align="center" valign="middle" >1.0</td><td align="center" valign="middle" >1.5</td><td align="center" valign="middle" >2.0</td><td align="center" valign="middle" >2.5</td><td align="center" valign="middle" >3.0</td></tr><tr><td align="center" valign="middle" >Rafa</td><td align="center" valign="middle" >194.0%%%(7.3)</td><td align="center" valign="middle" >188.3%%%(5.6)</td><td align="center" valign="middle" >187.2%%%(4.9)</td><td align="center" valign="middle" >179.4%%%(4.9)</td><td align="center" valign="middle" >172.0%%%(4.5)</td><td align="center" valign="middle" >163.8%%%(7.5)</td></tr><tr><td align="center" valign="middle" >Non-coated</td><td align="center" valign="middle" >134.4%%%(2.7)</td><td align="center" valign="middle" >111.6%%%(7.3)</td><td align="center" valign="middle" >104.5%%%(5.8)</td><td align="center" valign="middle" >86.1%%%(7.2)</td><td align="center" valign="middle" >72.2%%%(3.9)</td><td align="center" valign="middle" >56.6%%%(4.6)</td></tr></tbody></table></table-wrap><p>(5.5), at each depth of 0.5, 1.0, 1.5 and 2.0 cm respectively on the ultrasound images, whereas the intensity of the non-coated Tuohy needle was 134.8 (4.2), 116.8 (7.9), 94.5 (4.4), 67.6 (1.4) at each respective depth. <xref ref-type="table" rid="table2">Table 2</xref> was the intensity of the puncture angle set at 45 degrees and the intensity of Rafa was 194.0 (7.3), 188.3 (5.6), 187.2 (4.9), 179.2 (4.9), 172.0 (4.5), 163.8 (7.5) at each depth of 0.5, 1.0, 1.5, 2.0, 2.5 and 3.0 cm respectively on the ultrasound images, whereas the intensity of the non-coated Tuohy needle was 134.4 (2.7), 111.6 (7.3), 104.5 (5.8), 86.1 (7.2), 72.2 (3.9), 56.6 (4.6) at each respective depth.</p><p>In all the readings we took, the intensity of the Rafa was clearly higher than that of the non-coated needle (P &lt; 0.0001）.</p><p>In the test with the puncture angle set at 30 degrees, the intensity of the non-coated Tuohy needle decreased with the depth. On the other hand, the intensity of the Rafa did not decrease with the depth at all (<xref ref-type="fig" rid="fig1">Figure 1</xref>).</p><p>In the test with the puncture angle set at 45degrees, the intensity of the non-coated Tuohy needle decreased with the depth. On the other hand, the intensity of the Rafa decreased only slightly with the increase of depth (<xref ref-type="fig" rid="fig2">Figure 2</xref>).</p></sec><sec id="s5"><title>5. Discussions</title><p>We could prove the high visualization of the Rafa based on the results gotten for all the points where we gathered</p><fig id="fig1"  position="float"><label><xref ref-type="fig" rid="fig1">Figure 1</xref></label><caption><title> The intensity of each needle for the puncture angle of 30 degrees</title></caption><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/1-1920418x7.png"/></fig><fig id="fig2"  position="float"><label><xref ref-type="fig" rid="fig2">Figure 2</xref></label><caption><title> The intensity of each needle for the puncture angle of 45 degrees</title></caption><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/1-1920418x8.png"/></fig><p>data.</p><p>It has been generally reported that the larger the puncture angle is, the lower the visualization of the needle is [<xref ref-type="bibr" rid="scirp.68780-ref8">8</xref>] [<xref ref-type="bibr" rid="scirp.68780-ref9">9</xref>] . Therefore, the risk of the deep nerve block has arisen in comparison with the superficial nerve block [<xref ref-type="bibr" rid="scirp.68780-ref10">10</xref>] . This study also showed that the intensity of the non-coated Tuohy needle clearly decreased with the puncture angle and the depth. As the Rafa with sand-blasted steel at the tip of the bevel has shown validness on the ultrasound images regardless of the puncture angle and the depth, we are certain it will be effective for deep nerve blocks, such as the paravertebral nerve block and lumber plexus nerve block. It is also easy to check the tip position because only the tip is coated, not the entire needle.</p><p>The limitation of this study is caused by the lack of human study, due to the restrictions provided by the Blue Phantom, as it is not homogenous. Furthermore, we cannot measure the intensity at over 3 cm of depth because of the Blue Phantom. The puncture angles of this study are only 30 degrees and 45 degrees. In the future, we will need to perform this study in a clinical setting to confirm the clinical efficacy of this technique. As for the puncture angle, we will have to also perform the other puncture angles in next study.</p><p>In this study, the Rafa is a valid needle on the ultrasonic images. The Rafa will provide us the reliability and the safety to perform ultrasound-guided nerve blocks.</p></sec><sec id="s6"><title>Assistance with the Letter</title><p>None.</p></sec><sec id="s7"><title>Financial Support and Sponsorship</title><p>None.</p></sec><sec id="s8"><title>Conflict of Interest</title><p>None.</p></sec><sec id="s9"><title>Cite this paper</title><p>Yuki Ichikawa,Hironobu Ueshima,Hiroshi Otake,Akira Kitamura, (2016) Regardless of the Puncture Angle, a Tuohy Needle with Sand-Blasted Steel at the Tip of the Bevel Is a Valid Needle on the Ultrasonic Images. Open Journal of Anesthesiology,06,101-104. doi: 10.4236/ojanes.2016.67017</p></sec><sec id="s10"><title>NOTES</title></sec></body><back><ref-list><title>References</title><ref id="scirp.68780-ref1"><label>1</label><mixed-citation publication-type="other" xlink:type="simple">Vadi, M.G., Patel, N. and Stiegler, M.P. (2014) Local Anesthetic Systemic Toxicity after Combined Psoas Compartment-Sciatic Nerve Block: Analysis of Decision Factors and Diagnostic Delay. Anesthesiology, 120, 987-996. http://dx.doi.org/10.1097/ALN.0000000000000154</mixed-citation></ref><ref id="scirp.68780-ref2"><label>2</label><mixed-citation publication-type="other" xlink:type="simple">Jacob, A.K., Mantilla, C.B., Sviggum, H.P., Schroeder, D.R., Pagnano, M.W. and Hebl, J.R. (2011) Perioperative Nerve Injury after Total Hip Arthroplasty: Regional Anesthesia Risk during a 20-Year Cohort Study. Anesthesiology, 115, 1172-1178. http://dx.doi.org/10.1097/aln.0b013e3182326c20</mixed-citation></ref><ref id="scirp.68780-ref3"><label>3</label><mixed-citation publication-type="other" xlink:type="simple">Dhir, S., Ganapathy, S., Lindsay, P. and Athewal, G.S. (2007) Case Report: Ropivacaine Neurotoxicity at Clinical Doses in Interscalene Brachial Plexus Block. Canadian Journal of Anesthesia, 54, 912-916. http://dx.doi.org/10.1007/BF03026796</mixed-citation></ref><ref id="scirp.68780-ref4"><label>4</label><mixed-citation publication-type="other" xlink:type="simple">Ueshima, H. and Otake, H. (2016) A Tuohy Needle with Engrave Processing at the Tip of the Bevel Is the Safety Needle for Ultrasound-Guided Nerve Block. Journal of Clinical Anesthesia, 32, 82-83. http://dx.doi.org/10.1016/j.jclinane.2015.12.029</mixed-citation></ref><ref id="scirp.68780-ref5"><label>5</label><mixed-citation publication-type="other" xlink:type="simple">Ueshima, H., Nakagawa, M. and Otake, H. (2016) A Universal Echosupport Can Fix a Technique of Ultrasound-Guided Continuous Femoral Nerve Block. Journal of Clinical Anesthesia, 31, 191-192. http://dx.doi.org/10.1016/j.jclinane.2016.01.046</mixed-citation></ref><ref id="scirp.68780-ref6"><label>6</label><mixed-citation publication-type="other" xlink:type="simple">Ueshima, H. and Kitamura, A. (2015) A Needle with Sand-Blasted Steel at the Tip of the Bevel Is Valid for Ultrasound-Guided Deep Nerve Block. Journal of Clinical Anesthesia, 27, 363. http://dx.doi.org/10.1016/j.jclinane.2015.03.019</mixed-citation></ref><ref id="scirp.68780-ref7"><label>7</label><mixed-citation publication-type="other" xlink:type="simple">Ueshima, H. and Kitamura, A. (2015) The Use of a Needle Guide Kit Improves the Stability of Ultrasound-Guided Techniques. Journal of Anesthesia, 29, 803-804. http://dx.doi.org/10.1007/s00540-015-2021-0</mixed-citation></ref><ref id="scirp.68780-ref8"><label>8</label><mixed-citation publication-type="other" xlink:type="simple">Tsui, B.C.H. (2007) Facilitating Needle Alignment In-Plane to an Ultrasound Beam Using a Portable Laser Unit. Regional Anesthesia and Pain Medicine, 32, 84-88. http://dx.doi.org/10.1097/00115550-200701000-00015</mixed-citation></ref><ref id="scirp.68780-ref9"><label>9</label><mixed-citation publication-type="other" xlink:type="simple">Schafhater-Zoppoth, I., McCulloch, C.E. and Gray, A.T. (2004) Ultrasound Visibility of Needle Used for Regional Nerve Block: And in Vitro Study. Regional Anesthesia and Pain Medicine, 29, 480-488.</mixed-citation></ref><ref id="scirp.68780-ref10"><label>10</label><mixed-citation publication-type="other" xlink:type="simple">Klein, S.M., Faris, S., Celidonio, L., Tarantino, U., Fabbi, E. and Sabato, A.F. (2009) Retroperitoneal Haematoma in a Patient with Continuous Psoas Compartment Block and Enoxaparin Administration for Total Knee Replacement. British Journal of Anaesthesia, 103, 309-310. http://dx.doi.org/10.1093/bja/aep189</mixed-citation></ref></ref-list></back></article>