<?xml version="1.0" encoding="UTF-8"?><!DOCTYPE article  PUBLIC "-//NLM//DTD Journal Publishing DTD v3.0 20080202//EN" "http://dtd.nlm.nih.gov/publishing/3.0/journalpublishing3.dtd"><article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" dtd-version="3.0" xml:lang="en" article-type="research article"><front><journal-meta><journal-id journal-id-type="publisher-id">JMP</journal-id><journal-title-group><journal-title>Journal of Modern Physics</journal-title></journal-title-group><issn pub-type="epub">2153-1196</issn><publisher><publisher-name>Scientific Research Publishing</publisher-name></publisher></journal-meta><article-meta><article-id pub-id-type="doi">10.4236/jmp.2014.516162</article-id><article-id pub-id-type="publisher-id">JMP-50642</article-id><article-categories><subj-group subj-group-type="heading"><subject>Articles</subject></subj-group><subj-group subj-group-type="Discipline-v2"><subject>Physics&amp;Mathematics</subject></subj-group></article-categories><title-group><article-title>
 
 
  Dosimetric Evaluation of Heterogeneities in Small Circular Fields of 6 MV Photon Beams with EBT2 and EDR2 Films: Comparison with Monte Carlo Calculation
 
</article-title></title-group><contrib-group><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>li</surname><given-names>Mohammad</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Hassan</surname><given-names>A. Nedaie</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref><xref ref-type="corresp" rid="cor1"><sup>*</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Mehran</surname><given-names>YarAhmadi</given-names></name><xref ref-type="aff" rid="aff2"><sup>2</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Nooshin</surname><given-names>Banaee</given-names></name><xref ref-type="aff" rid="aff3"><sup>3</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Mansour</surname><given-names>Naderi</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Zohre</surname><given-names>Tizmaghz</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref></contrib></contrib-group><aff id="aff1"><addr-line>Radiotherapy Oncology Department, Cancer Research Centre, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran</addr-line></aff><aff id="aff2"><addr-line>Department of Medical Physics, Faculty of Medicine, Kurdistan University of Medical Sciences, Sanandaj, Iran</addr-line></aff><aff id="aff3"><addr-line>Young Researchers and Elites Club, Science and Research Branch, Islamic Azad University, Tehran, Iran</addr-line></aff><author-notes><corresp id="cor1">* E-mail:<email>Nedaieha@sina.tums.ac.ir(HAN)</email>;</corresp></author-notes><pub-date pub-type="epub"><day>17</day><month>10</month><year>2014</year></pub-date><volume>05</volume><issue>16</issue><fpage>1608</fpage><lpage>1616</lpage><history><date date-type="received"><day>2</day>	<month>August</month>	<year>2014</year></date><date date-type="rev-recd"><day>28</day>	<month>August</month>	<year>2014</year>	</date><date date-type="accepted"><day>15</day>	<month>September</month>	<year>2014</year></date></history><permissions><copyright-statement>&#169; Copyright  2014 by authors and Scientific Research Publishing Inc. </copyright-statement><copyright-year>2014</copyright-year><license><license-p>This work is licensed under the Creative Commons Attribution International License (CC BY). http://creativecommons.org/licenses/by/4.0/</license-p></license></permissions><abstract><p>
 
 
  The influence of dose variation due to heterogeneities in narrow photon beams used in stereotactic radiosurgery has been investigated. Since the lateral electronic disequilibrium and existence of steep dose gradients in small fields and the presence of heterogeneities can intensify these problems, in this study the effects of heterogeneities on 6 MV small photon beams produced by circular cone collimators with 5, 10, 15, 20 and 30 mm diameters are investigated. The heterogeneities include 3 cm Cork with density of 0.2 g/cm
  <sup>3</sup> instead of lung and 3 cm Polytetrafluoroethylene (P.T.F.E) with density of 2.2 g/cm
  <sup>3</sup> as bone. The measurements were carried out with EBT2 gafchromic and EDR2 radiographic films. Simulation was done by MCNP Monte Carlo Code (MCNP5). The depth dose curves in heterogeneous phantom were compared with homogeneous phantom. A good agreement was obtained within film and Monte Carlo calculations in presence of low density heterogeneity and also in the presence of high density heterogeneity. Monte Carlo results showed good agreement after stopping power correction.
 
</p></abstract><kwd-group><kwd>Stereotactic Radiosurgery</kwd><kwd> Heterogeneities</kwd><kwd> Film Dosimetry</kwd><kwd> Monte Carlo Calculation</kwd></kwd-group></article-meta></front><body><sec id="s1"><title>1. Introduction</title><p>In recent years, treatment by linear accelerator-based stereotactic radiosurgery (SRS) has increased significantly. This type of treatment modality employs narrowly-collimated, 6 - 10 radiation photon beams with field sizes smaller than 4 &#215; 4 cm<sup>2</sup> to limited volume of tumor. The aim of this technique is to deliver high dose with accurate position and dose of &#177;1 mm of &#177;5%, respectively [<xref ref-type="bibr" rid="scirp.50642-ref1">1</xref>] . To achive this accuracy, dosimetric measurements are of great importance. Because of the lateral electronic disequilibrium and steep dose gradient existence in large portion of these fields, dosimetry is more complicated [<xref ref-type="bibr" rid="scirp.50642-ref2">2</xref>] . Several studies have surveyed the dosimetry of narrow photon beams in a homogeneous medium [<xref ref-type="bibr" rid="scirp.50642-ref3">3</xref>] -[<xref ref-type="bibr" rid="scirp.50642-ref8">8</xref>] but the effect of density differences on such small beams has not been extensively studied [<xref ref-type="bibr" rid="scirp.50642-ref9">9</xref>] -[<xref ref-type="bibr" rid="scirp.50642-ref14">14</xref>] . In radiotherapy two types of tissue interface are important: low density heterogeneity/soft tissue, bone/soft tissue. The impact of heterogeneities may create perturbation in the dose distribution which can be deceptive in the exact localization of dose ranges. Clinically this may lead to the underdosing of the tumor and overdosing of the normal tissues. Many algorithms such as Batho and equivalent pathlenghth (EPL) avoid dealing with the complexity of charged particle transport (CPT) [<xref ref-type="bibr" rid="scirp.50642-ref15">15</xref>] . That is very important factor to estimate dose in regions without charged particle equilibrium (CPE) such as small fields of SRS in which heterogeneities can change electronic equilibrium. The majority of SRS treatment planning systems ignore the presence of heterogeneities due to computation time limitations and also assumption that multiple beam arrangements are compensating the perturbation effect. Because of the effect of heterogeneities on inaccuracy of the measurements, especially in small fields, selecting detectors with good spatial resolution, high accuracy and precision, dose rate and energy independence is very important. Although ionization chamber has good accuracy in measuring of absolute dose but in regions with steep dose gradient and charged particle disequilibrium, volume of detector affects on accuracy of the measurements [<xref ref-type="bibr" rid="scirp.50642-ref16">16</xref>] . Diode detectors are similar to ion chambers but their response is directionally dependant. TLD dosimeters also would require a large number of them to provide adequate information. Film is two-dimensional dosimeter with good spatial resolution for surveying pixel-by-pixel of measurements. Monte Carlo method provides a bridge between measurement and calculation and it is one of the best methods for situations in which physical measurements are difficult or impossible.</p><p>The aim of this study is to evaluate the effects of heterogeneities (low and high densities) of SRS fields irradiated by 6 MV photon beams using EBT2 and EDR2 films and also verification by Monte Carlo simulation.</p></sec><sec id="s2"><title>2. Materials and Methods</title><sec id="s2_1"><title>2.1. Phantoms</title><p>To evaluate dose perturbation caused by heterogeneities, PMMA slab phantom with 3 cm thickness of cork and density of 0.2 g/cm<sup>3</sup> (low density heterogeneity) and 3 cm thickness of Polytetrafluoroethylene (P.T.F.E) with density of 2.2 g/cm<sup>3</sup> (high density heterogeneity) were applied as lung and bone, respectively.</p></sec><sec id="s2_2"><title>2.2. Kodak EDR2 Film</title><p>The radiographic film type EDR2 (Kodak Extended Dose Range-2, Rochester, New York 14650) is a very slow speed and fine-grain film. Double emulsion active layers and 2.3 g/m<sup>2</sup> of cubic microcrystal AgBr grains with density of 6.53 g/cm<sup>3</sup> (average size of 0.24 micrometer) coated on 0.18 mm Easter base. EDR2 film is less sensitive to processor variations than XV2 film that allows processing of films in a routine film processor [<xref ref-type="bibr" rid="scirp.50642-ref17">17</xref>] . In order to prevent film batch differences, all films (with size of 35 &#215; 43 cm<sup>2</sup>) were chosen from the same batch [<xref ref-type="bibr" rid="scirp.50642-ref18">18</xref>] .</p></sec><sec id="s2_3"><title>2.3. Radiochromic Film</title><p>EBT2 Gafchromic films (International Specialty Products) were used as an alternative dosimeter with the composition of 40.85% H, 0.10% Li, 42.37% C, 0.01% N, 16.59% O, 0.04% Cl, 0.01% K and 0.01% Br, with effective atomic number of 6.84, as reported by manufacturer. EBT2 contains a yellow “marker” dye, which minimizes response differences caused by coating anomalies. EBT2 is nearly 10&#215; less sensitive to cool white fluorescent light than EBT films. The spectral absorbance peak of EBT2 is about 636 nm. According to the recommendation of the manufactures, all films should be scanned in landscape mode and red channel, because the yellow marker dye has no absorbance in the red portion of the spectrum and therefore has no effect on the film response in the red color channel. All films (size of 20 &#215; 25 cm<sup>2</sup>) were chosen from the same batch.</p></sec><sec id="s2_4"><title>2.4. Monte Carlo Simulation</title><p>The simulation was done by MCNP5 Monte Carlo (MC) radiation transport code (Los Alamos National Laboratory). The MCPLIB04 photon cross-section library was applied using data from ENDF/B-VI. The results from the MCNP5 calculations contain pulse height tally for energy or charge deposition, to calculate relative dose in this study.</p></sec><sec id="s2_5"><title>2.5. Dose Delivery</title><p>The Films were irradiated by 6 MV photon beams produced by Clinac 2100C/D (Varian Medical Systems, Inc., Palo Alto, CA) which were collimated by circular cone collimators with sizes of 5, 10, 15, 20, 30 mm diameters.</p></sec><sec id="s2_6"><title>2.6. Film Calibration</title><p>To acquire the relationship between the optical density of the EDR2 film and absorbed dose, The EDR2 film was placed perpendicular to the central beam axis under 5 cm of polystyrene with 15 cm of polystyrene underneath the film. The gantry and the collimator angles were 0˚ and the source-to-surface distance (SSD) was 100 cm. Eight different dose levels (25, 50, 75, 100, 125, 150, 200, 250 cGy) were delivered with a field size of 5 &#215; 5 cm<sup>2</sup>. An unexposed film was left to determine the optical density (OD) corresponding to the base and fog in such a calibration set (<xref ref-type="fig" rid="fig1">Figure 1</xref>(a)). For calibration of EBT2 film, the film was cut into 30 pieces with size of 2 &#215; 3 cm<sup>2</sup>, and the films were placed perpendicular to the central beam axis under 2 cm of polystyrene with 15 cm of polystyrene underneath the film, similar to EDR2 setup (<xref ref-type="fig" rid="fig1">Figure 1</xref>(b)). For each dose level, three pieces of films were chosen from different portions of original film. The irradiated films were scanned in RGB (Red, Green, Blue) mode and landscape direction 24 hours after irradiation. Before scanning the films, MICROTEK 9800 XL scanner was calibrated with Q-60E calibration strip.</p></sec><sec id="s2_7"><title>2.7. Measurement and Calculation</title><p>The shape of rectangular slab phantom is shown in <xref ref-type="fig" rid="fig2">Figure 2</xref>. For measurements, each film was packed tightly in the center of the phantom and the radiation beams were impinging parallel to the film. SSD was 100 cm and 100 MU was applied for each radiation beam at all irradiation steps of this study. In order to verify the simulations, calculated PDD and beam profiles of 10 &#215; 10 cm<sup>2</sup> field size were compared with those obtained by ionization chamber measurements. The same setup was repeated for circular cones as well. To reduce run time, electron and photon energy cut off were set to 0.5 and 0.01 MeV, respectively. Also one phase space file with histories of 3 billions was acquired after the flattening filter (the place that was constant for all of configurations). The dose</p><fig-group id="fig1"><label><xref ref-type="fig" rid="fig1">Figure 1</xref></label><caption><title> Calibration curves for (a) EDR2, (b) EBT2 films.</title></caption><fig id ="fig1_1"><label> (b)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x6.png"/></fig><fig id ="fig1_2"><label></label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x7.png"/></fig></fig-group><fig id="fig2"  position="float"><label><xref ref-type="fig" rid="fig2">Figure 2</xref></label><caption><title> Setup of examination contains rectangular slabs of polyethylene (soft tissue), 3 cm cork (low density heterogeneity) and 3 cm P.T.F.E ( high density heterogeneity)</title></caption><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x8.png"/></fig><p>voxels were different for each collimator sizes. The accepted statistical uncertainty was below 2% for all simulations.</p></sec></sec><sec id="s3"><title>3. Results and Discussion</title><sec id="s3_1"><title>3.1. Low Density Heterogeneity</title><p>The measured depth doses with EBT2 and EDR2 films in heterogeneous phantom included 3 cm cork along with those calculated by MCNP5 code are shown in <xref ref-type="fig" rid="fig3">Figure 3</xref>(a)-(d). Maximum dose variations for each field size are listed in <xref ref-type="table" rid="table1">Table 1</xref>. The results show dose reduction at central axis by depth increasing, this reduction is higher for smaller field sizes. Re-buildup region after the heterogeneity almost is independent of field size.</p></sec><sec id="s3_2"><title>3.2. High Density Heterogeneity</title><p>Like the low density heterogeneity, the measured and calculated data of heterogeneous phantom consisting of 3 cm P.T.F.E are shown in <xref ref-type="fig" rid="fig4">Figure 4</xref>. <xref ref-type="fig" rid="fig5">Figure 5</xref> illustrates the MCNP5 data after correction and also comparison between perturbations caused by heterogeneity. Maximum dose variations for each field size are tabulated in <xref ref-type="table" rid="table2">Table 2</xref>. As it can be seen (<xref ref-type="fig" rid="fig4">Figure 4</xref>), dose increases at the first layer on central axis, while MCNP5 shows dose increasing in central axis for 5, 10 mm collimator sizes and decreasing for 15, 20 and 30 mm collimator sizes.</p><p>As it was shown in <xref ref-type="fig" rid="fig4">Figure 4</xref>, in the presence of low density heterogeneity, significant perturbation can be observed inside the Cork that). For small field sizes surveyed in this study (5, 10, 15, 20 and 30 mm), lateral electronic disequilibrium is very noteworthy and this parameter increases inside the low density heterogeneity. Through the low density of Cork, secondary electrons that leave the field area will increase in comparison to water-equivalent media and this is more important parameter interns if forward range. Increasing the lateral scattering of secondary electrons causes dose reduction in central axis. Although the electron stopping power varies with logarithm of the mean excitation potential of each material, but the average of water-to-lung stopping power ratio is almost 1 [<xref ref-type="bibr" rid="scirp.50642-ref19">19</xref>] and this is consistence with results obtained in this study by Monte Carlo, EBT2 and EDR2 films inside the low density medium. Therefore dose variations in low density heterogeneity depend on density of cork. Also extent of decrease in dose at central axis depends on the field size of radiation beams. With decreasing field size, secondary electrons that leave the field area will increase and this will result in increasing downfall of central axis dose. After the second lung-tissue interface re-buildup is occurred due to decreased attenuation of photon in low density heterogeneity and consequently dose increases proportion to homogeneous phantom. For high density heterogeneity situation is the reverse. Both films showed increase at first layer of P.T.F.E because the films are different in physical density and electron density with bone. Film densities are near the soft tissue therefore, they show dose in unit density of films. Secondary electrons are also</p><fig-group id="fig3"><label><xref ref-type="fig" rid="fig3">Figure 3</xref></label><caption><title> Measured and calculated PDD in heterogeneous phantom with 3 cm of cork (low density) that was irradiated in (a) 5, (b) 10, (c) 15, (d) 20 and (e) 30 mm field sizes.</title></caption><fig id ="fig3_1"><label> (b)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x9.png"/></fig><fig id ="fig3_2"><label>(c)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x10.png"/></fig><fig id ="fig3_3"><label> (d)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x11.png"/></fig><fig id ="fig3_4"><label>(e)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x12.png"/></fig><fig id ="fig3_5"><label></label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x13.png"/></fig></fig-group><fig-group id="fig4"><label><xref ref-type="fig" rid="fig4">Figure 4</xref></label><caption><title> Measured and calculated PDD in heterogeneous phantom with 3 cm of P.T.F.E (high density) which was irradiated with 5, 10, 15, 20 and 30 mm diameter circular beams.</title></caption><fig id ="fig4_1"><label> (b)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x14.png"/></fig><fig id ="fig4_2"><label>(c)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x15.png"/></fig><fig id ="fig4_3"><label> (d)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x16.png"/></fig><fig id ="fig4_4"><label>(e)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x17.png"/></fig><fig id ="fig4_5"><label></label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x18.png"/></fig></fig-group><fig-group id="fig5"><label><xref ref-type="fig" rid="fig5">Figure 5</xref></label><caption><title> Calculated PDD in heterogeneous phantom before and after correction with 3 cm of P.T.F.E (high density) that irradiated with 5, 10, 15, 20 and 30 mm diameter circular beams.</title></caption><fig id ="fig5_1"><label> (b)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x19.png"/></fig><fig id ="fig5_2"><label>(c)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x20.png"/></fig><fig id ="fig5_3"><label> (d)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x21.png"/></fig><fig id ="fig5_4"><label>(e)</label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x22.png"/></fig><fig id ="fig5_5"><label></label><graphic mimetype="image"   position="float"  xlink:type="simple"  xlink:href="http://html.scirp.org/file/7-7501962x23.png"/></fig></fig-group><table-wrap id="table1" ><label><xref ref-type="table" rid="table1">Table 1</xref></label><caption><title> Maximum dose variations inside the cork relative to homogeneous phantom</title></caption><table><tbody><thead><tr><th align="center" valign="middle" >Field size (mm)</th><th align="center" valign="middle" >EDR2 (%)</th><th align="center" valign="middle" >EBT2 (%)</th><th align="center" valign="middle" >MCNP (%)</th></tr></thead><tr><td align="center" valign="middle" >5</td><td align="center" valign="middle" >43.35</td><td align="center" valign="middle" >35.76</td><td align="center" valign="middle" >40.1</td></tr><tr><td align="center" valign="middle" >10</td><td align="center" valign="middle" >36.75</td><td align="center" valign="middle" >32.45</td><td align="center" valign="middle" >37.2</td></tr><tr><td align="center" valign="middle" >15</td><td align="center" valign="middle" >31.26</td><td align="center" valign="middle" >28.14</td><td align="center" valign="middle" >30</td></tr><tr><td align="center" valign="middle" >20</td><td align="center" valign="middle" >23.56</td><td align="center" valign="middle" >20.82</td><td align="center" valign="middle" >25.7</td></tr><tr><td align="center" valign="middle" >30</td><td align="center" valign="middle" >16.40</td><td align="center" valign="middle" >13.35</td><td align="center" valign="middle" >17</td></tr></tbody></table></table-wrap><table-wrap id="table2" ><label><xref ref-type="table" rid="table2">Table 2</xref></label><caption><title> Maximum dose variations inside the P.T.F.E relative to homogeneous phantom</title></caption><table><tbody><thead><tr><th align="center" valign="middle" >Field size</th><th align="center" valign="middle" >EDR2 (%)</th><th align="center" valign="middle" >EBT2 (%)</th><th align="center" valign="middle" >MC (%)</th><th align="center" valign="middle" >Corrected MC (%)</th></tr></thead><tr><td align="center" valign="middle" >5 mm</td><td align="center" valign="middle" >7.16</td><td align="center" valign="middle" >6.12</td><td align="center" valign="middle" >4.74</td><td align="center" valign="middle" >7.3</td></tr><tr><td align="center" valign="middle" >10 mm</td><td align="center" valign="middle" >10.79</td><td align="center" valign="middle" >11.66</td><td align="center" valign="middle" >3.27</td><td align="center" valign="middle" >10.67</td></tr><tr><td align="center" valign="middle" >15 mm</td><td align="center" valign="middle" >5.9</td><td align="center" valign="middle" >6.65</td><td align="center" valign="middle" >−6.19</td><td align="center" valign="middle" >5.71</td></tr><tr><td align="center" valign="middle" >20 mm</td><td align="center" valign="middle" >5.27</td><td align="center" valign="middle" >6.9</td><td align="center" valign="middle" >−10.02</td><td align="center" valign="middle" >5.59</td></tr><tr><td align="center" valign="middle" >30 mm</td><td align="center" valign="middle" >6.34</td><td align="center" valign="middle" >6.1</td><td align="center" valign="middle" >−6.7</td><td align="center" valign="middle" >6.3</td></tr></tbody></table></table-wrap><p>deflected from multiple scattering and it is enhanced in higher atomic number materials like bone [<xref ref-type="bibr" rid="scirp.50642-ref20">20</xref>] which causes the change in energy and dose in central axis. Electron density in bone is smaller than soft tissue, therefore in Compton scattering area that depends to electron density, both EDR2 and EBT2 films show increase in dose.</p><p>Monte Carlo data show increasing of dose at first layer in central axis for 5, 10 mm collimator sizes because the effect of multiple scatterings on central axis is rule over and decreases the dose for 15, 20 and 30 mm collimator sizes. After high density heterogeneity, decrease of dose is observed because of shielding effect of bone. This effect is related to the thickness of bone and almost is independent of field size. Since base of Monte Carlo dose calculations typically report absorbed dose to the medium in which the particle transport took place [<xref ref-type="bibr" rid="scirp.50642-ref19">19</xref>] , comparison of Dose-to-medium (Dm) with dose-to-water (Dw), require a conversion of Dm to Dw. This conversion was made on the Bragg-Gray theory, Dw = Dm (Sw/Sm). Therefore as it was depicted in <xref ref-type="fig" rid="fig5">Figure 5</xref>, if the results obtained by Monte Carlo code are multiply to water-to-medium mass collision stopping power over the energy spectra of primary electrons at the point of interest [<xref ref-type="bibr" rid="scirp.50642-ref19">19</xref>] , MCNP data will appear near the data of the films.</p></sec></sec><sec id="s4"><title>4. Conclusion</title><p>Both EDR2 radiographic and EBT2 radiochromic films showed good agreement in low density medium as well as Monte Carlo Code. Results showed that dose variations in low density medium depend on density and field sizes that were used in this study (from 5 - 30 mm diameter). Inside the high density medium, the EDR2 and EBT2 results are similar but differ from MCNP data. After MCNP data correction using Bragg-Gray theory, both films and MCNP show good agreement with each other. Electron densities in high density medium and in Compton scattering area are important factors for variations and these variations are almost independent of field size.</p></sec><sec id="s5"><title>NOTES</title></sec></body><back><ref-list><title>References</title><ref id="scirp.50642-ref1"><label>1</label><mixed-citation publication-type="other" xlink:type="simple">Podgorsak, B., Pike, G.B., Pla, M., Olivier, A. and Souhami, L. (1990) Radiotherapy and Oncology, 17, 349-358. http://dx.doi.org/10.1016/0167-8140(90)90008-K</mixed-citation></ref><ref id="scirp.50642-ref2"><label>2</label><mixed-citation publication-type="other" xlink:type="simple">Wu, A., Zwicker, R.D., Kalend, A.M. and Zheng, Z. (1993) Medical Physics, 20, 777. http://dx.doi.org/10.1118/1.597032</mixed-citation></ref><ref id="scirp.50642-ref3"><label>3</label><mixed-citation publication-type="other" xlink:type="simple">Bjarngard, B.E., Tsai, J.S. and Rice, R.K. (1990) Medical Physics, 17, 794. http://dx.doi.org/10.1118/1.596475</mixed-citation></ref><ref id="scirp.50642-ref4"><label>4</label><mixed-citation publication-type="other" xlink:type="simple">Rice, R.K., Hansen, J.L., Svensson, G.K. and Siddon, R.L. (1987) Physics in Medicine and Biology, 32, 1087-1099. http://dx.doi.org/10.1088/0031-9155/32/9/002</mixed-citation></ref><ref id="scirp.50642-ref5"><label>5</label><mixed-citation publication-type="other" xlink:type="simple">Rustgi, SN. and Frye, D.M. (1995) Medical Physics, 22, 2117-2121. http://dx.doi.org/10.1118/1.597655</mixed-citation></ref><ref id="scirp.50642-ref6"><label>6</label><mixed-citation publication-type="other" xlink:type="simple">Westermark, M., Arndt, J., Nilsson, B. and Brahme, A. (2000) Physics in Medicine and Biology, 45, 685-702. http://dx.doi.org/10.1088/0031-9155/45/3/308</mixed-citation></ref><ref id="scirp.50642-ref7"><label>7</label><mixed-citation publication-type="other" xlink:type="simple">Martens, C., De Wagter, C. and De Neve, W. (2000) Physics in Medicine and Biology, 45, 2519-2530. http://dx.doi.org/10.1088/0031-9155/45/9/306</mixed-citation></ref><ref id="scirp.50642-ref8"><label>8</label><mixed-citation publication-type="other" xlink:type="simple">Verhaegen, F., Das, I.J. and Palmans, H. (1998) Physics in Medicine and Biology, 43, 2755-2768. http://dx.doi.org/10.1088/0031-9155/43/10/006</mixed-citation></ref><ref id="scirp.50642-ref9"><label>9</label><mixed-citation publication-type="other" xlink:type="simple">Rustgi, S.N., Rustgi, A.K., Jiang, S.B. and Ayyangar, KM. (1998) Physics in Medicine and Biology, 43, 3509-3518. http://dx.doi.org/10.1088/0031-9155/43/12/009</mixed-citation></ref><ref id="scirp.50642-ref10"><label>10</label><mixed-citation publication-type="other" xlink:type="simple">Ahn, Y.C., Lee, K.C., Kim, D.Y., Huh, S.J., Yeo, I.H., Lim, D.H., Kim, M.K., Shin, K.H., Park, S. and Chang, S.H. (2000) International Journal of Radiation Oncology · Biology · Physics, 48, 501-505. http://dx.doi.org/10.1016/S0360-3016(00)00612-X</mixed-citation></ref><ref id="scirp.50642-ref11"><label>11</label><mixed-citation publication-type="other" xlink:type="simple">Cheung, J.Y., Yu, K.N., Yu, C.P. and Ho, R.T. (2001) Applied Radiation and Isotopes, 54, 461-465. http://dx.doi.org/10.1016/S0969-8043(00)00283-9</mixed-citation></ref><ref id="scirp.50642-ref12"><label>12</label><mixed-citation publication-type="other" xlink:type="simple">Jones, A.O., Das, I.J. and Jones Jr., F.L. (2003) Medical Physics, 30, 296-300. http://dx.doi.org/10.1118/1.1539040</mixed-citation></ref><ref id="scirp.50642-ref13"><label>13</label><mixed-citation publication-type="other" xlink:type="simple">Carrasco, P., Jornet, N., Duch, M.A., Panettieri, V., Weber, L., Eudaldo, T., Ginjaume, M. and Ribas, M. (2007) Medical Physics, 34, 3323-3333. http://dx.doi.org/10.1118/1.2750972</mixed-citation></ref><ref id="scirp.50642-ref14"><label>14</label><mixed-citation publication-type="other" xlink:type="simple">Klein, E.E., Chin, L.M., Rice, R.K. and Mijnheer, B.J. (1993) International Journal of Radiation Oncology, Biology, Physics, 27, 419-427.</mixed-citation></ref><ref id="scirp.50642-ref15"><label>15</label><mixed-citation publication-type="other" xlink:type="simple">Jones, A.O. and Das, I.J. (2005) Medical Physics, 32, 766-776. http://dx.doi.org/10.1118/1.1861154</mixed-citation></ref><ref id="scirp.50642-ref16"><label>16</label><mixed-citation publication-type="other" xlink:type="simple">Pappas, E., Maris, T.G., Zacharopoulou, F., Papadakis, A., Manolopoulos, S., Green, S. and Wojnecki, C. (2008) Medical Physics, 35, 4640-4648. http://dx.doi.org/10.1118/1.2977829</mixed-citation></ref><ref id="scirp.50642-ref17"><label>17</label><mixed-citation publication-type="other" xlink:type="simple">Dogan, N., Leybovich, L.B. and Sethi, A. (2002) Physics in Medicine and Biology, 47, 4121-4130. http://dx.doi.org/10.1088/0031-9155/47/22/314</mixed-citation></ref><ref id="scirp.50642-ref18"><label>18</label><mixed-citation publication-type="other" xlink:type="simple">Shi, C., Papanikolaou, N., Yan, Y., Weng, X. and Jiang, H. (2006) Journal of Applied Clinical Medical Physics, 7, 1-8.</mixed-citation></ref><ref id="scirp.50642-ref19"><label>19</label><mixed-citation publication-type="other" xlink:type="simple">Siebers, J.V., Keall, P.V., Nahum, A.E. and Mohan, R. (2000) Physics in Medicine and Biology, 45, 983-995. http://dx.doi.org/10.1088/0031-9155/45/4/313</mixed-citation></ref><ref id="scirp.50642-ref20"><label>20</label><mixed-citation publication-type="other" xlink:type="simple">AAPM Report (2004) Task Group 65. Tissue Inhomogeneity Corrections for Megavoltage Photon Beams. The American Association of Physicists in Medicine, No. 85, Madison.</mixed-citation></ref></ref-list></back></article>