<?xml version="1.0" encoding="UTF-8"?><!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v3.0 20080202//EN" "http://dtd.nlm.nih.gov/publishing/3.0/journalpublishing3.dtd">
<article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" dtd-version="3.0" xml:lang="en" article-type="research article">
 <front>
  <journal-meta>
   <journal-id journal-id-type="publisher-id">
    wjnst
   </journal-id>
   <journal-title-group>
    <journal-title>
     World Journal of Nuclear Science and Technology
    </journal-title>
   </journal-title-group>
   <issn pub-type="epub">
    2161-6795
   </issn>
   <issn publication-format="print">
    2161-6809
   </issn>
   <publisher>
    <publisher-name>
     Scientific Research Publishing
    </publisher-name>
   </publisher>
  </journal-meta>
  <article-meta>
   <article-id pub-id-type="doi">
    10.4236/wjnst.2025.151001
   </article-id>
   <article-id pub-id-type="publisher-id">
    wjnst-138738
   </article-id>
   <article-categories>
    <subj-group subj-group-type="heading">
     <subject>
      Articles
     </subject>
    </subj-group>
    <subj-group subj-group-type="Discipline-v2">
     <subject>
      Engineering, Physics 
     </subject>
     <subject>
       Mathematics
     </subject>
    </subj-group>
   </article-categories>
   <title-group>
    Assessment of Optimal Use, Maintenance, Repair and Calibration of Radiation Monitoring Instruments in Nigeria
   </title-group>
   <contrib-group>
    <contrib contrib-type="author" xlink:type="simple">
     <name name-style="western">
      <surname>
       Samuel Mofolorunsho
      </surname>
      <given-names>
       Oyeyemi
      </given-names>
     </name> 
     <xref ref-type="aff" rid="aff1"> 
      <sup>1</sup>
     </xref>
    </contrib>
    <contrib contrib-type="author" xlink:type="simple">
     <name name-style="western">
      <surname>
       Olumide Olaife
      </surname>
      <given-names>
       Akerele
      </given-names>
     </name> 
     <xref ref-type="aff" rid="aff1"> 
      <sup>1</sup>
     </xref>
    </contrib>
    <contrib contrib-type="author" xlink:type="simple">
     <name name-style="western">
      <surname>
       Sunday Ufuoma
      </surname>
      <given-names>
       Obarhua
      </given-names>
     </name> 
     <xref ref-type="aff" rid="aff1"> 
      <sup>1</sup>
     </xref>
    </contrib>
    <contrib contrib-type="author" xlink:type="simple">
     <name name-style="western">
      <surname>
       Francis Adole
      </surname>
      <given-names>
       Agada
      </given-names>
     </name> 
     <xref ref-type="aff" rid="aff1"> 
      <sup>1</sup>
     </xref>
    </contrib>
    <contrib contrib-type="author" xlink:type="simple">
     <name name-style="western">
      <surname>
       Wasiu Kofoworola
      </surname>
      <given-names>
       Ayuba
      </given-names>
     </name> 
     <xref ref-type="aff" rid="aff1"> 
      <sup>1</sup>
     </xref>
    </contrib>
    <contrib contrib-type="author" xlink:type="simple">
     <name name-style="western">
      <surname>
       Helen Enikpi
      </surname>
      <given-names>
       Alakiu
      </given-names>
     </name> 
     <xref ref-type="aff" rid="aff1"> 
      <sup>1</sup>
     </xref>
    </contrib>
    <contrib contrib-type="author" xlink:type="simple">
     <name name-style="western">
      <surname>
       David Olakanmi
      </surname>
      <given-names>
       Olaniyi
      </given-names>
     </name> 
     <xref ref-type="aff" rid="aff1"> 
      <sup>1</sup>
     </xref>
    </contrib>
    <contrib contrib-type="author" xlink:type="simple">
     <name name-style="western">
      <surname>
       Ethel Ebere
      </surname>
      <given-names>
       Ofoegbu
      </given-names>
     </name> 
     <xref ref-type="aff" rid="aff2"> 
      <sup>2</sup>
     </xref>
    </contrib>
   </contrib-group> 
   <aff id="aff1">
    <addr-line>
     aNational Institute of Radiation Protection and Research, University of Ibadan, Ibadan, Oyo State, Nigeria
    </addr-line> 
   </aff> 
   <aff id="aff2">
    <addr-line>
     aNigerian Nuclear Regulatory Authority, Abuja, Nigeria
    </addr-line> 
   </aff> 
   <pub-date pub-type="epub">
    <day>
     06
    </day> 
    <month>
     01
    </month>
    <year>
     2025
    </year>
   </pub-date> 
   <volume>
    15
   </volume> 
   <issue>
    01
   </issue>
   <fpage>
    1
   </fpage>
   <lpage>
    16
   </lpage>
   <history>
    <date date-type="received">
     <day>
      8,
     </day>
     <month>
      October
     </month>
     <year>
      2024
     </year>
    </date>
    <date date-type="published">
     <day>
      3,
     </day>
     <month>
      October
     </month>
     <year>
      2024
     </year> 
    </date> 
    <date date-type="accepted">
     <day>
      3,
     </day>
     <month>
      January
     </month>
     <year>
      2025
     </year> 
    </date>
   </history>
   <permissions>
    <copyright-statement>
     © Copyright 2014 by authors and Scientific Research Publishing Inc. 
    </copyright-statement>
    <copyright-year>
     2014
    </copyright-year>
    <license>
     <license-p>
      This work is licensed under the Creative Commons Attribution International License (CC BY). http://creativecommons.org/licenses/by/4.0/
     </license-p>
    </license>
   </permissions>
   <abstract>
    The response and performance of radiation detectors for accurate measurements and effective use for radiological safety in medical, industrial, and nuclear sectors are based on the optimal use, maintenance, repair and calibration of radiation monitoring instruments in a secondary standard dosimetry laboratory. In Nigeria, the suboptimal performances of these instruments are attributed to inadequate maintenance practices, insufficient calibration, and limited awareness of proper equipment handling for optimal use. This study assesses the current practices related to the optimal use, maintenance, repair, and calibration of radiation detection equipment across Nigeria’s six geopolitical zones. Using a cross-sectional survey approach, data were collected from Ninety (90) radiation monitoring equipment operators, Radiation Safety Officers, and frontline responders to evaluate their knowledge, awareness, and practices concerning equipment usage, operation, storage, handling, and calibration. The findings reveal significant gaps in knowledge of usage (trained is 43.2%, not trained is 56.8%) and inconsistencies in maintenance practices (as indicated by the regression analysis (β = 0.51, p &lt; 0.01), particularly regarding specialized instruments such as the PackEye, Mobile Detection System (MDS), Radionuclide Identifinder (RID), and Personal Radiation Detectors (PRD). While there is high awareness of the need for regular calibration and handling training, the lack of standardized protocols and training alignment poses challenges to the effective use of these instruments. This study underscores the importance of comprehensive training programs, standardized maintenance protocols, and enhanced awareness initiatives to optimize the usage, performance and safety of radiation monitoring instruments in Nigeria.
   </abstract>
   <kwd-group> 
    <kwd>
     Radiation Monitoring Instruments
    </kwd> 
    <kwd>
      Detectors
    </kwd> 
    <kwd>
      Calibration
    </kwd> 
    <kwd>
      Radiation Safety Officers
    </kwd> 
    <kwd>
      PackEye
    </kwd> 
    <kwd>
      Radionuclide Identifinder
    </kwd> 
    <kwd>
      Personal Radiation Detectors
    </kwd>
   </kwd-group>
  </article-meta>
 </front>
 <body>
  <sec id="s1">
   <title>1. Introduction</title>
   <p>Radiation monitoring instruments are essential tools for ensuring radiological safety and nuclear security in various practices, including medical, industrial, and nuclear energy sectors. These instruments, which consist of a detector medium, an electronic component, and a display unit, are designed to detect and measure radiation levels to prevent unnecessary exposure to ionizing radiation. Accurate knowledge of the usage and functioning of these instruments is critical for radiation safety; however, improper maintenance, poor storage, inadequate calibration, and poor handling can result in equipment failure, leading to inaccurate measurements, unexpected downtime, and potential radiation overexposure to workers and the public <xref ref-type="bibr" rid="scirp.138738-1">
     [1]
    </xref> <xref ref-type="bibr" rid="scirp.138738-2">
     [2]
    </xref>.</p>
   <p>The significance of maintaining radiation monitoring equipment cannot be overstated. Research indicates that many failures of radiation monitoring instruments are attributed to factors such as circuit board issues, wear and tear, improper use, manufacturing defects, and harsh environmental conditions <xref ref-type="bibr" rid="scirp.138738-3">
     [3]
    </xref> <xref ref-type="bibr" rid="scirp.138738-4">
     [4]
    </xref>. Equipment failure can also lead to substantial financial losses due to repair costs and the need for replacement, as well as safety risks in critical scenarios <xref ref-type="bibr" rid="scirp.138738-5">
     [5]
    </xref>. Studies have emphasized the importance of robust maintenance strategies and periodic calibration to mitigate these risks <xref ref-type="bibr" rid="scirp.138738-6">
     [6]
    </xref> <xref ref-type="bibr" rid="scirp.138738-7">
     [7]
    </xref>. However, in Nigeria, a lack of standardized maintenance practices and limited awareness of equipment handling among frontline officers and operators have contributed to the suboptimal performance of these instruments <xref ref-type="bibr" rid="scirp.138738-8">
     [8]
    </xref>.</p>
   <p>The International Atomic Energy Agency (IAEA) initiated the Coordinated Research Project J02014, which aimed at advancing the maintenance, repair, and calibration of radiation detection equipment globally, especially those deployed at the border and other strategic points to prevent illicit trafficking of nuclear and radioactive material. The assessment of performance, optimization, maintenance, repair, and calibration activities for radiation monitoring instruments mostly used for nuclear monitoring and security is very paramount. The users of equipment such as the PackEye, Mobile Detection System (MDS), Radionuclide Identifinder (RID), and Personal Radiation Detectors (PRD) in Nigeria, despite their widespread use, have a significant gap in knowledge regarding their usage, maintenance and calibration, which affects their response, reliability and effectiveness <xref ref-type="bibr" rid="scirp.138738-2">
     [2]
    </xref>. There is a major problem in ensuring the equipment performs to specification and expectation for a long period of time, if the users lack the theoretical knowledge of the operation of the detector</p>
   <p>Previous studies have highlighted the importance of regular calibration and maintenance of radiation detection instruments to ensure accuracy and functionality <xref ref-type="bibr" rid="scirp.138738-9">
     [9]
    </xref> <xref ref-type="bibr" rid="scirp.138738-10">
     [10]
    </xref>. Tamura <xref ref-type="bibr" rid="scirp.138738-11">
     [11]
    </xref> constructed a POMDP model for maintenance action to analyse the properties of the optimal maintenance policy. However, a comprehensive understanding of the maintenance strategies and training needs among users in developing countries like Nigeria is lacking. The Secondary Standard Dosimetry Laboratory located at the University of Ibadan was established as a Technical Support Organization of the Nigerian Nuclear Regulatory Authority. It has traceability to IAEA primary Lab located at Siberdorf in Vienna, Austria. It ensures standardization of radiation measuring instruments in Nigeria through calibration using Cesium and X-ray source. The Institute has competent personnel who are experienced in the use and calibration of radiation-measuring equipment.</p>
   <p>This paper presents the results of a survey conducted among radiation equipment users in Nigeria to evaluate their awareness and understanding of equipment operation, storage, handling, and calibration. The findings reveal critical gaps in knowledge and practices, underscoring the need for comprehensive training programs and standardized maintenance protocols to ensure the effective use of radiation monitoring instruments in the country.</p>
  </sec><sec id="s2">
   <title>2. Materials and Methods</title>
   <p>A cross-sectional survey design was employed to evaluate the maintenance, repair, and calibration practices of radiation monitoring instruments across Nigeria’s six geopolitical zones (South-South, South-East, South-West, North-Central, North-East, and North-West). The study targeted users of various radiation detection equipment mostly used for nuclear monitoring and nuclear security, including the PackEye, Mobile Detection System (MDS), Radionuclide Identifinder (RID), and Personal Radiation Detectors (PRD) as shown in <xref ref-type="table" rid="table1">
     Table 1
    </xref>. This design was chosen to capture the experiences and practices across diverse locations and operational environments.</p>
   <table-wrap id="table1">
    <label>
     <xref ref-type="table" rid="table1">
      Table 1
     </xref></label>
    <caption>
     <title>
      <xref ref-type="bibr" rid="scirp.138738-"></xref>Table 1. Equipment specifications and functions.</title>
    </caption>
    <table class="MsoTableGrid custom-table" border="0" cellspacing="0" cellpadding="0"> 
     <tr> 
      <td class="custom-bottom-td acenter" width="6.16%"><p style="text-align:center">S/N</p></td> 
      <td class="custom-bottom-td acenter" width="17.33%"><p style="text-align:center">Equipment Name</p></td> 
      <td class="custom-bottom-td acenter" width="15.31%"><p style="text-align:center">Detection Medium</p></td> 
      <td class="custom-bottom-td acenter" width="15.61%"><p style="text-align:center">Sensitivity</p></td> 
      <td class="custom-bottom-td acenter" width="17.65%"><p style="text-align:center">Measuring Range</p></td> 
      <td class="custom-bottom-td acenter" width="13.67%"><p style="text-align:center">Energy Range</p></td> 
      <td class="custom-bottom-td acenter" width="14.26%"><p style="text-align:center">Radiation Type</p></td> 
     </tr> 
     <tr> 
      <td class="custom-top-td acenter" width="6.16%"><p style="text-align:center">1</p></td> 
      <td class="custom-top-td acenter" width="17.33%"><p style="text-align:center">Rapiscan Mobile Detection System (MDS)</p></td> 
      <td class="custom-top-td acenter" width="15.31%"><p style="text-align:center">Organic plastic scintillator (Gamma), He-3 tube (Neutron)</p></td> 
      <td class="custom-top-td acenter" width="15.61%"><p style="text-align:center">See below the table</p></td> 
      <td class="custom-top-td acenter" width="17.65%"><p style="text-align:center"></p></td> 
      <td class="custom-top-td acenter" width="13.67%"><p style="text-align:center"></p></td> 
      <td class="custom-top-td acenter" width="14.26%"><p style="text-align:center">Gamma, Neutron</p></td> 
     </tr> 
     <tr> 
      <td class="acenter" width="6.16%"><p style="text-align:center">2</p></td> 
      <td class="acenter" width="17.33%"><p style="text-align:center">Thermo Scientific Radeye PRD</p></td> 
      <td class="acenter" width="15.31%"><p style="text-align:center">NaI (Tl) scintillation detectors</p></td> 
      <td class="acenter" width="15.61%"><p style="text-align:center">150 cps/µSv/hr (Cs-137);</p><p style="text-align:center">3000 cps/µSv/hr (60 keV);</p></td> 
      <td class="acenter" width="17.65%"><p style="text-align:center">0 to 250 µSv/hr;</p><p style="text-align:center">0 to 800 kcps</p></td> 
      <td class="acenter" width="13.67%"><p style="text-align:center">30 keV to 1.3 MeV</p></td> 
      <td class="acenter" width="14.26%"><p style="text-align:center">Gamma</p></td> 
     </tr> 
     <tr> 
      <td class="acenter" width="6.16%"><p style="text-align:center">3</p></td> 
      <td class="acenter" width="17.33%"><p style="text-align:center">FLIR Radionuclide Identifinder (RID)</p></td> 
      <td class="acenter" width="15.31%"><p style="text-align:center">Nal; Geiger-Müller tube</p></td> 
      <td class="acenter" width="15.61%"><p style="text-align:center"></p></td> 
      <td class="acenter" width="17.65%"><p style="text-align:center">0 μSv/h to 500 μSv/h (NaI); 100 μSv/h - 1 Sv/h (Geiger-Muller)</p></td> 
      <td class="acenter" width="13.67%"><p style="text-align:center">20 keV to 3 MeV</p></td> 
      <td class="acenter" width="14.26%"><p style="text-align:center">Gamma</p></td> 
     </tr> 
     <tr> 
      <td class="acenter" width="6.16%"><p style="text-align:center">4</p></td> 
      <td class="acenter" width="17.33%"><p style="text-align:center">PackEYE BackPack</p></td> 
      <td class="acenter" width="15.31%"><p style="text-align:center">Plastic scintillator (Gamma); Li-6 doped flat scintillation detectors (Neutron)</p></td> 
      <td class="acenter" width="15.61%"><p style="text-align:center">Greater than 3000 cps/microSv/h at 662 keV for gamma</p></td> 
      <td class="acenter" width="17.65%"><p style="text-align:center">0.05 to 26 µSv/hr (Gamma);</p><p style="text-align:center">0.5 to 256 cps (neutron)</p></td> 
      <td class="acenter" width="13.67%"><p style="text-align:center">20 keV to 3 MeV (Gamma)</p></td> 
      <td class="acenter" width="14.26%"><p style="text-align:center">Gamma, Neutron</p></td> 
     </tr> 
    </table>
   </table-wrap>
   <p>Sensitivity of MDS: 1) Gamma: Will detect 1000 g of 235U (HEU) and 10 g of 239 Pu in 20 uR/hr background at a passage speed of 5 mph (8 km/h), at a distance of 39 in (1 m); 2) Neutron: Will detect less than 200 g of plutonium in a shielded container that reduces the gamma flux to 1% of the unshielded gamma flux.</p>
   <sec id="s2_1">
    <title>2.1. Data Collection</title>
    <p>Data were collected using a semi-structured questionnaire administered to Ninety (90) radiation monitoring equipment operators, safety officers, and frontline workers across the six geopolitical zones (See Appendix I). The questionnaire, designed to capture both quantitative and qualitative data, was divided into four sections:</p>
    <p>1) Demographic Information Collected: Basic demographic data such as age, gender, education level, and geographical location.</p>
    <p>2) Awareness of Equipment Operation and Handling: Assessed respondents’ awareness and understanding of the operation and handling of radiation detection instruments.</p>
    <p>3) Knowledge of Equipment Specifications and Functions: Gathered information on the technical knowledge of the equipment, including sensitivity, detection range, and calibration intervals.</p>
    <p>4) General Knowledge and Maintenance Practices: Explored overall knowledge of maintenance practices, storage conditions, calibration procedures, and common faults.</p>
    <p>Data collection occurred from March to May 2024, with face-to-face interviews conducted to clarify responses and gather in-depth qualitative insights.</p>
   </sec>
   <sec id="s2_2">
    <title>2.2. Data Analysis</title>
    <p>The quantitative data were analyzed using SPSS version 26.0, employing descriptive statistics (frequencies, percentages, means, standard deviations) to summarize demographic characteristics and levels of awareness, knowledge, and maintenance practices. Inferential statistics, including chi-square tests, were used to examine associations between demographic variables and knowledge or awareness levels. Cross-tabulation identified patterns in maintenance practices across different zones.</p>
    <p>Qualitative data from open-ended questions and interviews were analyzed thematically, with responses coded to identify emerging themes related to maintenance challenges, training needs, and optimization strategies.</p>
   </sec>
  </sec><sec id="s3">
   <title>3. Results</title>
   <sec id="s3_1">
    <title>3.1. Demographic Data</title>
    <p>Results of the evaluation of the demographic data revealed that there were 10% (9) female and 90% (81) male from the 90 respondents. The range of their ages in percentage was represented in <xref ref-type="fig" rid="fig1">
      Figure 1
     </xref> (pie chart) below. Responses were obtained from different locations in the six geopolitical zones across the country as shown in <xref ref-type="table" rid="table2">
      Table 2
     </xref>.</p>
    <fig id="fig1" position="float">
     <label>Figure 1</label>
     <caption>
      <title>Figure 1. Percentage for age ranges.</title>
     </caption>
     <graphic mimetype="image" position="float" xlink:type="simple" xlink:href="https://html.scirp.org/file/1090546-rId14.jpeg?20250106033604" />
    </fig>
    <table-wrap id="table2">
     <label>
      <xref ref-type="table" rid="table2">
       Table 2
      </xref></label>
     <caption>
      <title>
       <xref ref-type="bibr" rid="scirp.138738-"></xref>Table 2. Operational locations.</title>
     </caption>
     <table class="MsoTableGrid custom-table" border="0" cellspacing="0" cellpadding="0"> 
      <tr> 
       <td class="custom-bottom-td acenter" width="25.22%"><p style="text-align:center">REGION</p></td> 
       <td class="custom-bottom-td acenter" width="27.79%"><p style="text-align:center">OPERATIONAL LOCATIONS</p></td> 
       <td class="custom-bottom-td acenter" width="23.50%"><p style="text-align:center">POPULATION</p></td> 
       <td class="custom-bottom-td acenter" width="23.49%"><p style="text-align:center">% POPULATION</p></td> 
      </tr> 
      <tr> 
       <td class="custom-top-td acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="custom-top-td acenter" width="27.79%"><p style="text-align:center">Port Harcourt</p></td> 
       <td class="custom-top-td acenter" width="23.50%"><p style="text-align:center">16</p></td> 
       <td class="custom-top-td acenter" width="23.49%"><p style="text-align:center">17.9</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center">SOUTH-SOUTH</p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Bonny</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">4</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">4.5</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Delta</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">16</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">17.9</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Edo</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">3</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">3.4</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center">SOUTH-WEST</p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Lagos</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">13</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">14.6</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Ogun</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">4</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">4.5</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Ibadan</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">5</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">5.6</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Ekiti</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">2</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">2.3</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center">NORTH-CENTRAL</p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Abuja</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">8</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">8.9</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Plateau</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">2</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">2.3</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center">NORTH-WEST</p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Kano</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">3</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">3.4</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Katsina/Kaduna</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">4</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">4.5</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center"></p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Sokoto</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">2</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">2.3</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center">NORTH-EAST</p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Adamawa/Yobe/</p><p style="text-align:center">Maiduguri/Gombe</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">4</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">4.5</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="25.22%"><p style="text-align:center">SOUTH-EAST</p></td> 
       <td class="acenter" width="27.79%"><p style="text-align:center">Enugu</p></td> 
       <td class="acenter" width="23.50%"><p style="text-align:center">3</p></td> 
       <td class="acenter" width="23.49%"><p style="text-align:center">3.4</p></td> 
      </tr> 
     </table>
    </table-wrap>
   </sec>
   <sec id="s3_2">
    <title>3.2. Awareness of Equipment Operation and Handling</title>
    <p>This session detailed the respondent’s awareness of the equipment operation, calibration procedure, training needs, and handling responsibilities.</p>
    <fig id="fig2" position="float">
     <label>Figure 2</label>
     <caption>
      <title>Figure 2. % Awareness of equipment operation.</title>
     </caption>
     <graphic mimetype="image" position="float" xlink:type="simple" xlink:href="https://html.scirp.org/file/1090546-rId15.jpeg?20250106033604" />
    </fig>
    <p>The bar chart in <xref ref-type="fig" rid="fig2">
      Figure 2
     </xref> presents the respondents’ awareness of the equipment operations. It shows that 88.9% of users are aware of the operation of PRD while 1.1% of the users can operate a multipurpose detector.</p>
    <fig id="fig3" position="float">
     <label>Figure 3</label>
     <caption>
      <title>Figure 3. % Awareness of equipment handling.</title>
     </caption>
     <graphic mimetype="image" position="float" xlink:type="simple" xlink:href="https://html.scirp.org/file/1090546-rId16.jpeg?20250106033605" />
    </fig>
    <p>From <xref ref-type="fig" rid="fig3">
      Figure 3
     </xref>, averagely 89% of users have an awareness of handling and use of detectors through regular training, which was made compulsory for everybody, while 83% believed that handling requires some specialized training. Less than 20% have not handled the detector at all.</p>
    <p>
     <xref ref-type="fig" rid="fig4">
      Figure 4
     </xref> presents the type and level of training acquired by the respondent, which shows that 64% of users had basic knowledge of training, 23% had intermediate and 13% had advanced training on the use of the equipment. While 43% had undergone basic training, only about 30% had undergone intermediate or advance training.</p>
    <fig id="fig4" position="float">
     <label>Figure 4</label>
     <caption>
      <title>Figure 4. % Type and level of training acquired.</title>
     </caption>
     <graphic mimetype="image" position="float" xlink:type="simple" xlink:href="https://html.scirp.org/file/1090546-rId17.jpeg?20250106033605" />
    </fig>
    <p>
     <xref ref-type="fig" rid="fig5">
      Figure 5
     </xref> presents general awareness of radiation equipment while <xref ref-type="fig" rid="fig6">
      Figure 6
     </xref> summarized knowledge of equipment specifications and functions.</p>
    <p>It is shown in <xref ref-type="fig" rid="fig5">
      Figure 5
     </xref> that 86% of the users had a general awareness of the equipment with descriptions, while only 14% had a general awareness with examples.</p>
    <p>Respondents showed good knowledge of the technical specifications and detection capabilities of these devices, particularly the RID (87% agreement) as represented in <xref ref-type="fig" rid="fig6">
      Figure 6
     </xref>.</p>
    <fig id="fig5" position="float">
     <label>Figure 5</label>
     <caption>
      <title>Figure 5. General awareness of radiation equipment with description and examples.</title>
     </caption>
     <graphic mimetype="image" position="float" xlink:type="simple" xlink:href="https://html.scirp.org/file/1090546-rId18.jpeg?20250106033605" />
    </fig>
    <p>
     <xref ref-type="table" rid="table3">
      Table 3
     </xref> presents the awareness of the respondents as to who should handle radiation equipment, and it shows that Radiation Safety Officers (RSOs) are viewed as the most qualified personnel to handle radiation equipment (40%).</p>
    <p>
     <xref ref-type="table" rid="table4">
      Table 4
     </xref> presents the awareness of equipment storage knowledge, and it shows that 35.5% are of the view that the equipment should be stored in a warehouse, a safe and secure place.</p>
    <fig id="fig6" position="float">
     <label>Figure 6</label>
     <caption>
      <title>Figure 6. % Summary of knowledge of equipment specifications and functions.</title>
     </caption>
     <graphic mimetype="image" position="float" xlink:type="simple" xlink:href="https://html.scirp.org/file/1090546-rId19.jpeg?20250106033605" />
    </fig>
    <table-wrap id="table3">
     <label>
      <xref ref-type="table" rid="table3">
       Table 3
      </xref></label>
     <caption>
      <title>
       <xref ref-type="bibr" rid="scirp.138738-"></xref>Table 3. Awareness of who should handle radiation equipment.</title>
     </caption>
     <table class="MsoTableGrid custom-table" border="0" cellspacing="0" cellpadding="0"> 
      <tr> 
       <td class="custom-bottom-td acenter" width="10.27%"><p style="text-align:center">S/N</p></td> 
       <td class="custom-bottom-td acenter" width="40.40%"><p style="text-align:center">PERSONNEL THAT SHOULD HANDLE THE EQUIPMENT</p></td> 
       <td class="custom-bottom-td acenter" width="22.02%"><p style="text-align:center">RESPONDENT</p></td> 
       <td class="custom-bottom-td acenter" width="27.31%"><p style="text-align:center">% RESPONDENT</p></td> 
      </tr> 
      <tr> 
       <td class="custom-top-td acenter" width="10.27%"><p style="text-align:center">1</p></td> 
       <td class="custom-top-td aleft" width="40.40%"><p style="text-align:left">Radiation Safety Officer (RSO)</p></td> 
       <td class="custom-top-td acenter" width="22.02%"><p style="text-align:center">36</p></td> 
       <td class="custom-top-td acenter" width="27.31%"><p style="text-align:center">40</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="10.27%"><p style="text-align:center">2</p></td> 
       <td class="aleft" width="40.40%"><p style="text-align:left">Trained and Certified Personnel</p></td> 
       <td class="acenter" width="22.02%"><p style="text-align:center">26</p></td> 
       <td class="acenter" width="27.31%"><p style="text-align:center">28.9</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="10.27%"><p style="text-align:center">3</p></td> 
       <td class="aleft" width="40.40%"><p style="text-align:left">Classified Radiation Workers</p></td> 
       <td class="acenter" width="22.02%"><p style="text-align:center">15</p></td> 
       <td class="acenter" width="27.31%"><p style="text-align:center">16.7</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="10.27%"><p style="text-align:center">4</p></td> 
       <td class="aleft" width="40.40%"><p style="text-align:left">Operators/Radiographers</p></td> 
       <td class="acenter" width="22.02%"><p style="text-align:center">8</p></td> 
       <td class="acenter" width="27.31%"><p style="text-align:center">8.9</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="10.27%"><p style="text-align:center">5</p></td> 
       <td class="aleft" width="40.40%"><p style="text-align:left">Medical Physicists</p></td> 
       <td class="acenter" width="22.02%"><p style="text-align:center">5</p></td> 
       <td class="acenter" width="27.31%"><p style="text-align:center">5.5</p></td> 
      </tr> 
     </table>
    </table-wrap>
    <table-wrap id="table4">
     <label>
      <xref ref-type="table" rid="table4">
       Table 4
      </xref></label>
     <caption>
      <title>
       <xref ref-type="bibr" rid="scirp.138738-"></xref>Table 4. Awareness of equipment storage in the facility.</title>
     </caption>
     <table class="MsoTableGrid custom-table" border="0" cellspacing="0" cellpadding="0"> 
      <tr> 
       <td class="custom-bottom-td acenter" width="12.41%"><p style="text-align:center">S/N</p></td> 
       <td class="custom-bottom-td acenter" width="39.66%"><p style="text-align:center">LOCATION FOR EQUIPMENT STORAGE</p></td> 
       <td class="custom-bottom-td acenter" width="21.40%"><p style="text-align:center">RESPONDENT</p></td> 
       <td class="custom-bottom-td acenter" width="26.53%"><p style="text-align:center">% RESPONDENT</p></td> 
      </tr> 
      <tr> 
       <td class="custom-top-td acenter" width="12.41%"><p style="text-align:center">1</p></td> 
       <td class="custom-top-td aleft" width="39.66%"><p style="text-align:left">In the Instrumentation room</p></td> 
       <td class="custom-top-td acenter" width="21.40%"><p style="text-align:center">2</p></td> 
       <td class="custom-top-td acenter" width="26.53%"><p style="text-align:center">2.2</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="12.41%"><p style="text-align:center">2</p></td> 
       <td class="aleft" width="39.66%"><p style="text-align:left">In a warehouse, safe and secured place</p></td> 
       <td class="acenter" width="21.40%"><p style="text-align:center">32</p></td> 
       <td class="acenter" width="26.53%"><p style="text-align:center">35.5</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="12.41%"><p style="text-align:center">3</p></td> 
       <td class="aleft" width="39.66%"><p style="text-align:left">In the equipment box/case</p></td> 
       <td class="acenter" width="21.40%"><p style="text-align:center">5</p></td> 
       <td class="acenter" width="26.53%"><p style="text-align:center">5.6</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="12.41%"><p style="text-align:center">4</p></td> 
       <td class="aleft" width="39.66%"><p style="text-align:left">In a Bunker, well shielded area</p></td> 
       <td class="acenter" width="21.40%"><p style="text-align:center">25</p></td> 
       <td class="acenter" width="26.53%"><p style="text-align:center">27.8</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="12.41%"><p style="text-align:center">5</p></td> 
       <td class="aleft" width="39.66%"><p style="text-align:left">In a Controlled Area</p></td> 
       <td class="acenter" width="21.40%"><p style="text-align:center">21</p></td> 
       <td class="acenter" width="26.53%"><p style="text-align:center">23.3</p></td> 
      </tr> 
      <tr> 
       <td class="acenter" width="12.41%"><p style="text-align:center">6</p></td> 
       <td class="aleft" width="39.66%"><p style="text-align:left">In the office of RSO/Medical Physicists</p></td> 
       <td class="acenter" width="21.40%"><p style="text-align:center">5</p></td> 
       <td class="acenter" width="26.53%"><p style="text-align:center">5.6</p></td> 
      </tr> 
     </table>
    </table-wrap>
    <p>
     <xref ref-type="table" rid="table5">
      Table 5
     </xref> presents the awareness of procedure for calibration, which shows that awareness of calibration procedures is moderate (69.1%), but the remaining 30.9% either lack awareness or report no procedures.</p>
    <p>The questionnaire also assesses the Knowledge of Equipment Specifications and Functions for MDS, PACKEYE, RID and PRD. (See Appendix I)</p>
   </sec>
  </sec><sec id="s4">
   <title>4. Discussions</title>
   <p>The demographic data highlight a significant gender imbalance (90% male) and</p>
   <table-wrap id="table5">
    <label>
     <xref ref-type="table" rid="table5">
      Table 5
     </xref></label>
    <caption>
     <title>
      <xref ref-type="bibr" rid="scirp.138738-"></xref>Table 5. Awareness of procedure for calibration of equipment.</title>
    </caption>
    <table class="MsoTableGrid custom-table" border="0" cellspacing="0" cellpadding="0"> 
     <tr> 
      <td class="custom-bottom-td acenter" width="76.83%"><p style="text-align:center"></p></td> 
      <td class="custom-bottom-td acenter" width="11.58%"><p style="text-align:center">YES</p></td> 
      <td class="custom-bottom-td acenter" width="11.59%"><p style="text-align:center">NO</p></td> 
     </tr> 
     <tr> 
      <td class="custom-top-td acenter" width="76.83%"><p style="text-align:center">Is there a procedure in place for regular calibration of equipment?</p></td> 
      <td class="custom-top-td acenter" width="11.58%"><p style="text-align:center">56</p></td> 
      <td class="custom-top-td acenter" width="11.59%"><p style="text-align:center">26</p></td> 
     </tr> 
     <tr> 
      <td class="acenter" width="76.83%"><p style="text-align:center">%</p></td> 
      <td class="acenter" width="11.58%"><p style="text-align:center">69.1</p></td> 
      <td class="acenter" width="11.59%"><p style="text-align:center">30.9</p></td> 
     </tr> 
    </table>
   </table-wrap>
   <p>an aging workforce predominantly aged 45 and above, indicating the need for more inclusive recruitment strategies and succession planning. Geographically, respondents are concentrated in the South-South and South-West regions of Nigeria, revealing gaps in operational reach and awareness in the North-East and North-West. This uneven distribution suggests the need for targeted efforts to ensure uniform knowledge and capability across all regions.</p>
   <p>Awareness levels vary significantly across different radiation detection equipment. While the Personal Radiation Detector (PRD) shows high awareness (88.9%), specialized equipment like the Multipurpose detector (1.1%), as represented in <xref ref-type="fig" rid="fig2">
     Figure 2
    </xref>, has lower awareness, indicating a need for more focused equipment training. The majority of respondents (83.3%), as indicated in <xref ref-type="fig" rid="fig3">
     Figure 3
    </xref>, have experience handling radiation equipment, aligning with global best practices that emphasize rigorous training. However, some respondents (16.7%) lack this experience, highlighting operational access constraints.</p>
   <p>From <xref ref-type="fig" rid="fig4">
     Figure 4
    </xref>, the result reveals that there are notable gaps between the type and level of training required versus the training received with only 43.2% of respondents having received intermediate or advanced training necessary for the safe and effective handling of radiation detection equipment. Regression analysis reveals that training is a significant predictor of proper maintenance practices (β = 0.51, p &lt; 0.01). This finding underscores the potential of targeted training interventions to enhance equipment maintenance and reduce failures. While prior studies have emphasized the importance of training <xref ref-type="bibr" rid="scirp.138738-7">
     [7]
    </xref>, this study provides quantitative evidence of the direct impact of advanced training on maintenance outcomes.</p>
   <p>There is a stark contrast between general awareness (86% with descriptions) and awareness with specific examples (14%), as shown in <xref ref-type="fig" rid="fig5">
     Figure 5
    </xref>, suggesting that theoretical knowledge does not always translate into practical understanding. This gap highlights the importance of practical, example-based training to ensure effective practice in radiation detection. The training programs should include verification of equipment, calibration and advanced handling techniques. This insight is particularly relevant for regulatory bodies like the Nigerian Nuclear Regulatory Authority (NNRA) and international agencies like the IAEA, which could use these findings to design more effective training curricula tailored to the specific needs of operators in Nigeria and similar settings. Radiation Safety Officers (RSOs) are viewed as the most qualified personnel to handle radiation detection equipment (40%), as represented in <xref ref-type="table" rid="table3">
     Table 3
    </xref>, but there is a need to diversify training to include other classified personnel that might be involved in the use of these detectors. This broader approach would enhance operational flexibility and resilience.</p>
   <p>Diverse opinions on storage practices suggest a lack of standardized information and protocols, which could lead to safety risks or inefficiencies. Comparing these findings with the literature, the low awareness of proper storage practices is particularly concerning. According to <xref ref-type="bibr" rid="scirp.138738-9">
     [9]
    </xref>, improper storage of radiation detection instruments can lead to equipment degradation, inaccurate readings, and increased exposure risks. This study corroborates that equipment storage remains a critical yet overlooked component of radiation safety management. Similarly, as represented in <xref ref-type="table" rid="table5">
     Table 5
    </xref>, awareness of calibration procedures is moderate (69.1%), but the remaining 30.9% either lack awareness or report no procedures, indicating a need for consistent policies and training to ensure equipment reliability.</p>
   <p>The knowledge of equipment specifications and functions, such as the Mobile Detection System (MDS), PackEye, Radionuclide Identifinder (RID), and Personal Radiation Detector (PRD), was also relatively high. Respondents showed good knowledge of the technical specifications and detection capabilities of these devices, particularly the RID (87% agreement), as represented in <xref ref-type="fig" rid="fig6">
     Figure 6
    </xref>. This aligns with the findings of <xref ref-type="bibr" rid="scirp.138738-10">
     [10]
    </xref> who reported that understanding the functionality of radiation detection instruments is essential for effective deployment and usage. The strong correlation between knowledge of equipment specifications and effective maintenance practices (r = 0.62, p &lt; 0.01) highlights the need for enhanced training programs that focus on both technical knowledge and practical application. This study provides new insights into the importance of technical knowledge as a determinant of proper maintenance and safety protocols. This connection suggests that increasing technical knowledge could enhance maintenance practices, reduce equipment failures, and improve overall radiation safety.</p>
  </sec><sec id="s5">
   <title>5. Recommendations</title>
   <p>1) There is a need to bridge the critical gap in knowledge of usage, training, and standardized procedures.</p>
   <p>2) Standardized maintenance protocols should be developed by the competent Authority and significant disparities in training needs versus training received should be itemized.</p>
   <p>3) Training programs should be focused on specific equipment types, such as MDS, PACKEYE, and RID. This is essential to bridge these existing gaps between frontline responders and radiation safety officers.</p>
   <p>4) Purposely built storage facilities should be made available at every facility of operators and calibration practices across different regions and facilities are crucial for minimizing safety risks and enhancing operational efficiency.</p>
   <p>5) There is a need for the establishment of consistent guidelines for storage handling.</p>
   <p>6) Implementation of region-specific and role-specific awareness programs to ensure uniform knowledge and skills distribution among all relevant personnel, especially in underrepresented regions, should be considered, and policies that mandate standardized training and certification for all users of radiation detection equipment, supported by regular audits and compliance checks should be implemented.</p>
   <p>7) Management of operators and users of radiation measuring equipment should have an inclusive workforce Planning and Diversity that promotes gender and age diversity in recruitment and training strategies to build a more sustainable and effective radiation safety management system.</p>
  </sec><sec id="s6">
   <title>6. Conclusion</title>
   <p>This study has revealed that users of radiation measuring equipment in Nigeria lack the necessary understanding of the storage and maintenance of the equipment with which they are familiar. It also shows that there is an existing gap in practical training for optimal performance. It provides valuable insights into the performance, maintenance, repair, and calibration practices of radiation monitoring instruments in Nigeria, highlighting critical gaps in knowledge, training, and standardized procedures. The data indicate that while there is a general understanding of the importance of regular calibration and handling training, there is a lack of standardized maintenance protocols and significant disparities in training needs versus training received. By addressing the above recommendations, the reliability, accuracy, and safety of radiation monitoring instruments would be improved, thereby enhancing radiological protection and safety management across Nigeria.</p>
  </sec><sec id="s7">
   <title>Acknowledgements</title>
   <p>We sincerely acknowledge the support of the International Atomic Energy Agency (IAEA) through the Coordinated Research Project (Jo2014), which played a crucial role in funding this study. The financial assistance, resources, and expert guidance provided by the IAEA have greatly facilitated our research efforts. We also express our deep gratitude to the team at the Secondary Standard Dosimetry Laboratory, Nigeria, whose expertise and dedication were instrumental in the successful completion of this research.</p>
  </sec><sec id="s8">
   <title>Appendix I: Research Questionnaire</title>
   <p>CRP J02014: ADVANCING MAINTENANCE, REPAIR AND CALIBRATION OF RADIATION DETECTION EQUIPMENT</p>
   <p>RESEARCH CONTRACT TITLED “PERFORMANCE ASSESSMENT AND OPTIMIZATION OF MAINTENANCE, REPAIR AND CALIBRATION ACTIVITIES FOR RADIATION MONITORING INSTRUMENTS IN NIGERIA.”</p>
   <p>This questionnaire is specially designed for research purposes; as such, the information supplied will be treated as confidential. Please feel free to give honest information as it applies to you.</p>
   <p>Thank you for your co-operation,</p>
   <p>INSTRUCTION:</p>
   <p>Kindly tick (√) where necessary and complete with possible answers where applicable.</p>
   <p>SECTION A (DEMOGRAPHIC DATA)</p>
   <p>SECTION B (AWARENESS OF EQUIPMENT OPERATION AND HANDLING)</p>
   <p>SECTION C (KNOWLEDGE OF EQUIPMENT SPECIFICATION AND FUNCTIONS)</p>
   <p>Kindly tick (√) where it applies to you</p>
   <p>SD: STRONGLY DISAGREE</p>
   <p>D: DISAGREE</p>
   <p>A: AGREE</p>
   <p>SA: STRONGLY AGREE</p>
   <table class="MsoTableGrid custom-table" border="0" cellspacing="0" cellpadding="0"> 
    <tr> 
     <td class="custom-bottom-td acenter" width="5.78%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td acenter" width="67.59%"><p style="text-align:center">MOBILE DETECTION SYSTEM (MDS) SCALE</p></td> 
     <td class="custom-bottom-td acenter" width="6.65%"><p style="text-align:center">SD</p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center">D</p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center">A</p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center">SA</p></td> 
    </tr> 
    <tr> 
     <td class="custom-top-td acenter" width="5.78%"><p style="text-align:center">22</p></td> 
     <td class="custom-top-td aleft" width="67.59%"><p style="text-align:left">I am aware MDS is a tool to survey large contaminated area</p></td> 
     <td class="custom-top-td acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">23</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">MDS involves using air to discover localized hot spots and map the event</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">24</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It can easily be deployed in any type of moving vehicle to search for unusual or abnormal radioactivity</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">25</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It comprised a large volume, highly sensitive organic plastic scintillator detector</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">26</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">The data is stored in a .dbf file, which can be easily exported to other programs</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">27</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">MDS uses MAPTRACK, a standalone software program that helps load any type of standard commercial or military maps</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">28</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">MDS is an ideal search and discover tool that may be utilized by nearly anyone with very little training required</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">29</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">The system contains a battery that when fully charged</p><p style="text-align:left">provides 8 hours of continuous use</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">30</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">External power capability is present with both AC and DC</p><p style="text-align:left">Adapters</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">31</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">The MDS has been designed for civilian and military responders</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">32</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It is the mobile detection solution for those who need fast, dependable threat assessment for tactical operations involving radiation</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">33</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It is a proactive surveillance which measures and also searches out and locates stolen/lost radiation sources</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">34</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">The MDS has been successfully tested and adopted by other military personnel for use on their FOX reconnaissance vehicles</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="custom-bottom-td acenter" width="5.78%"><p style="text-align:center">35</p></td> 
     <td class="custom-bottom-td aleft" width="67.59%"><p style="text-align:left">The MDS has also successfully been deployed in helicopters, emergency vehicles and by secret service agencies</p></td> 
     <td class="custom-bottom-td acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="custom-bottom-td custom-top-td acenter" width="5.78%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="67.59%"><p style="text-align:center">PACKEYE SCALE</p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="6.65%"><p style="text-align:center">SD</p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="6.66%"><p style="text-align:center">D</p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="6.66%"><p style="text-align:center">SA</p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="6.66%"><p style="text-align:center">A</p></td> 
    </tr> 
    <tr> 
     <td class="custom-top-td acenter" width="5.78%"><p style="text-align:center">36</p></td> 
     <td class="custom-top-td aleft" width="67.59%"><p style="text-align:left">PackEye is necessary to detect radiological contamination</p></td> 
     <td class="custom-top-td acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">37</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">PackEye has nothing to do with radiation of any kind</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">38</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It helps locate gamma emitting radioactive source in large areas rapidly</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">39</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Fast detection and locating of radioactive sources via high sensitivity plastic detector</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">40</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It ensures maximum sensitivity through automatic adjustment to background variations</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">41</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Its user-friendly interface with simple LED level, alarm and status indication including NBR feedback (red and green)</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">42</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It makes use of unchallenged light weight of 7.5 kg</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">43</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">The power consumption of PackEye is low</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">44</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Its operation time 30 hr (20 hr with Bluetooth<sup>TM</sup> data transfer to optional PDA)</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">45</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Prone to error by handlers</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="custom-bottom-td acenter" width="5.78%"><p style="text-align:center">46</p></td> 
     <td class="custom-bottom-td aleft" width="67.59%"><p style="text-align:left">Both the professional and non-professional can handle it</p></td> 
     <td class="custom-bottom-td acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="custom-bottom-td custom-top-td acenter" width="5.78%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="67.59%"><p style="text-align:center">RADIATION IDENTIFINDERS</p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-bottom-td custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="custom-top-td acenter" width="5.78%"><p style="text-align:center">47</p></td> 
     <td class="custom-top-td aleft" width="67.59%"><p style="text-align:left">They quickly locate and measure radioactive sources with confidence</p></td> 
     <td class="custom-top-td acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="custom-top-td acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">48</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">They deliver rapid radioisotope identification to reduce time to action and expedite response measures</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">49</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">They offer increased sensitivity, flexible power management, and industry-leading communication features</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">50</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It is a secondary verification tool to deliver the spectroscopic information needed to identify a specific nuclide</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">51</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">They deliver rapid isotope identification against the ANSI N42.34 library for threat determination</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">52</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">They are a great choice for security teams to distinguish between true threats and benign sources without disrupting the flow of an event where there is large gathering</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">53</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It helps to separate innocent sources such as medical personnel from potential threats</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">54</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Multi-layer approach is required for physical security to be effective</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center"></p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Personal Radiation Detectors (PRD) Scale</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center">SD</p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center">D</p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center">SA</p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center">A</p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">55</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Its small electronic devices are used to detect the illicit transport of radioactive materials</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">56</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It is also known as radiation pagers</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">57</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">They are used for screening during patrols or at events</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">58</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Most PRDs provide a digital display of the exposure rate and a visual, auditory, or vibratory alarm at preset thresholds</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">59</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">Designed to be clipped to the wearer’s belt, weigh less than a pound, and use battery power</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">60</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">They generally employ scintillation technology to detect gamma radiation</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">61</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">The devices have separate indicators to show if neutrons are detected</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">62</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">PRDs alert the wearer of the presence of radioactive material</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">68</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It can be used to interdict the illicit movement of radioactive material</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">69</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">PRDs are one of several complementary radiation measurement devices used in interdiction missions</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
    <tr> 
     <td class="acenter" width="5.78%"><p style="text-align:center">70</p></td> 
     <td class="aleft" width="67.59%"><p style="text-align:left">It makes use of scintillators containing lithium to indicate the presence of neutron radiation</p></td> 
     <td class="acenter" width="6.65%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
     <td class="acenter" width="6.66%"><p style="text-align:center"></p></td> 
    </tr> 
   </table>
   <p>SECTION D: GENERAL KNOWLEDGE</p>
   <p>Discussion Guide</p>
  </sec>
 </body><back>
  <ref-list>
   <title>References</title>
   <ref id="scirp.138738-ref1">
    <label>1</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Adjei, D., Darko, E.O., Annkah, J.K., Amoako, J.K., Ofori, K., Emi-Reynolds, G., et al. (2013) Analysis of Calibration Results of Radiation Survey Meters Used for Area Monitoring. Radiation Protection Dosimetry, 156, 506-513. &gt;https://doi.org/10.1093/rpd/nct087 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref2">
    <label>2</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Pibida, L., Estes, B., Mejias, M. and Klemic, G. (2021) Recalibration Intervals for Radiation Detection Instruments. &gt;https://nvlpubs.nist.gov/nistpubs/TechnicalNotes/NIST.TN.2146.pdf 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref3">
    <label>3</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     International Atomic Energy Agency (2000) Calibration of Radiation Protection Monitoring Instruments, Safety Reports Series No. 16.&gt;https://www.iaea.org/ns/coordinet 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref4">
    <label>4</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Al-Sultan, K.S. and Duffuaa, S.O. (1995) Maintenance Control via Mathematical Programming. Journal of Quality in Maintenance Engineering, 1, 36-46. &gt;https://doi.org/10.1108/13552519510096341 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref5">
    <label>5</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Ben-Daya, M. and Duffuaa, S.O. (1995) Maintenance and Quality: The Missing Link. Journal of Quality in Maintenance Engineering, 1, 20-26. &gt;https://doi.org/10.1108/13552519510083110 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref6">
    <label>6</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Liyanage, J.P. and Kumar, U. (2003) Towards a Value-Based View on Operations and Maintenance Performance Management. Journal of Quality in Maintenance Engineering, 9, 333-350. &gt;https://doi.org/10.1108/13552510310503213 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref7">
    <label>7</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Velmurugan, R.S. and Dhingra, T. (2015) Maintenance Strategy Selection and Its Impact in Maintenance Function. International Journal of Operations&amp;Production Management, 35, 1622-1661. &gt;https://doi.org/10.1108/ijopm-01-2014-0028 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref8">
    <label>8</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Pintelon, L. and Parodi-Herz, A. (2008) Maintenance: An Evolutionary Perspective. In: Springer Series in Reliability Engineering, Springer, 21-48. &gt;https://doi.org/10.1007/978-1-84800-011-7_2 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref9">
    <label>9</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     McDonald, J.C. (2004) Calibration Measurements and Standards for Radiation Protection Dosimetry. Radiation Protection Dosimetry, 109, 317-321. &gt;https://doi.org/10.1093/rpd/nch308 
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref10">
    <label>10</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Owusu-Banahene, J., Darko, E.O., Appiah, P. and Owusu-Manteaw, P. (2018) Calibration of Survey Meters at the Secondary Standards Dosimetry Laboratory at the Radiation Protection Institute of Ghana Tomic Energy Commission. International Journal of Science and Research, 8, 1386-1391.
    </mixed-citation>
   </ref>
   <ref id="scirp.138738-ref11">
    <label>11</label>
    <mixed-citation publication-type="other" xlink:type="simple">
     Tamura, N. (2023) Analysis of a POMDP Model for an Optimal Maintenance Problem with Multiple Imperfect Repairs. American Journal of Operations Research, 13, 133-146. &gt;https://doi.org/10.4236/ajor.2023.136008
    </mixed-citation>
   </ref>
  </ref-list>
 </back>
</article>