<?xml version="1.0" encoding="UTF-8"?><!DOCTYPE article  PUBLIC "-//NLM//DTD Journal Publishing DTD v3.0 20080202//EN" "http://dtd.nlm.nih.gov/publishing/3.0/journalpublishing3.dtd"><article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" dtd-version="3.0" xml:lang="en" article-type="research article"><front><journal-meta><journal-id journal-id-type="publisher-id">OJST</journal-id><journal-title-group><journal-title>Open Journal of Stomatology</journal-title></journal-title-group><issn pub-type="epub">2160-8709</issn><publisher><publisher-name>Scientific Research Publishing</publisher-name></publisher></journal-meta><article-meta><article-id pub-id-type="doi">10.4236/ojst.2020.1010029</article-id><article-id pub-id-type="publisher-id">OJST-103879</article-id><article-categories><subj-group subj-group-type="heading"><subject>Articles</subject></subj-group><subj-group subj-group-type="Discipline-v2"><subject>Medicine&amp;Healthcare</subject></subj-group></article-categories><title-group><article-title>
 
 
  Comparative Dimensional and Microscopic Analysis of the Strength and Hardness of Internal Fixation Plates
 
</article-title></title-group><contrib-group><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Victor</surname><given-names>Tieghi Neto</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Eduardo</surname><given-names>Stedile Fiamoncini</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Matheus</surname><given-names>Falsarelli Ferreira</given-names></name><xref ref-type="aff" rid="aff2"><sup>2</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Danilo</surname><given-names>Luiz Flumignan</given-names></name><xref ref-type="aff" rid="aff3"><sup>3</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Luis</surname><given-names>Geraldo Vaz</given-names></name><xref ref-type="aff" rid="aff4"><sup>4</sup></xref></contrib><contrib contrib-type="author" xlink:type="simple"><name name-style="western"><surname>Eduardo</surname><given-names>Sanches Gonçales</given-names></name><xref ref-type="aff" rid="aff1"><sup>1</sup></xref><xref ref-type="aff" rid="aff2"><sup>2</sup></xref><xref ref-type="corresp" rid="cor1"><sup>*</sup></xref></contrib></contrib-group><aff id="aff4"><addr-line>Department of Dental and Prosthetic Materials, Araraquara School of Dentistry, University of Sao Paulo, Araraquara, Brazil</addr-line></aff><aff id="aff3"><addr-line>Instituto Federal de Educa&amp;amp;#231;&amp;amp;#227;o, Ciência e Tecnologia de S&amp;amp;#227;o Paulo, IFSP, Mat&amp;amp;#227;o, Brazil</addr-line></aff><aff id="aff1"><addr-line>Department of Stomatology, Bauru School of Dentistry, University of Sao Paulo, Bauru, Brazil</addr-line></aff><aff id="aff2"><addr-line>Bauru School of Dentistry, University of Sao Paulo, Bauru, Brazil</addr-line></aff><pub-date pub-type="epub"><day>12</day><month>10</month><year>2020</year></pub-date><volume>10</volume><issue>10</issue><fpage>310</fpage><lpage>320</lpage><history><date date-type="received"><day>20,</day>	<month>August</month>	<year>2020</year></date><date date-type="rev-recd"><day>27,</day>	<month>October</month>	<year>2020</year>	</date><date date-type="accepted"><day>30,</day>	<month>October</month>	<year>2020</year></date></history><permissions><copyright-statement>&#169; Copyright  2014 by authors and Scientific Research Publishing Inc. </copyright-statement><copyright-year>2014</copyright-year><license><license-p>This work is licensed under the Creative Commons Attribution International License (CC BY). http://creativecommons.org/licenses/by/4.0/</license-p></license></permissions><abstract><p>
 
 
  The fixation stability achieved with the use of plates and screws in oral and maxillofacial surgery is a decisive factor in treatment success. The mechanical and structural properties of the internal fixation materials have direct influence on the dimensional stability and resistance of a fixation system, thus influencing treatment outcomes. This study proposed to analyze the dimensional and resistance patterns of titanium plates used for obtaining stable fix
  ation in orthognathic surgery and craniofacial trauma. For this study, 30 conventional 2.0
   
  mm straight four hole plates with bridge from three brands commercialized in Brazil, were subjected to macroscopic, microscopic, strength and hardness analysis. The dimensional measurements were performed using a digital caliper. Energy-dispersive X-ray spectroscopy analysis was performed by scanning electron microscopy to analyze the chemical composition of the samples. The mechanical resistance tests were performed with a universal testing machine. The samples were then submitted to Vickers hardness analysis, complying with the standards of ASTM E92. The data collected from the dimensional study was submitted to statistical analysis of the coefficient of variation, while the values obtained during the mechanical tests were analyzed by variance (ANOVA) and Tukey’s test (p &lt;
   
  0.05). The sample groups presented different performances in resistance, hardness, size and surface, even though they were reported to be chemically similar compounds that allowed us to conclude the plates from Group 2 were more resistant than groups 1 and 3.
 
</p></abstract><kwd-group><kwd>Fracture Fixation Internal</kwd><kwd> Biocompatible Materials</kwd><kwd> Compressive Strength</kwd></kwd-group></article-meta></front><body><sec id="s1"><title>1. Introduction</title><p>The treatment of skeletal deformities has undergone major advances in recent years. In the early days of buccomaxillofacial surgery and traumatology, the treatment of patients affected by these diseases was accomplished using fixation with steel wire, dental attachments, gutters, and orthopedic techniques, associated with the use of the intermaxillary block to favor the immobilization of the bone fragments [<xref ref-type="bibr" rid="scirp.103879-ref1">1</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref2">2</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref3">3</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref4">4</xref>].</p><p>With the advancement of the Weston-synthesis techniques, several types of materials were developed for fixation of mandibular fractures and osteotomies, and the use of plates and screws has been the current standard treatment [<xref ref-type="bibr" rid="scirp.103879-ref2">2</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref3">3</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref5">5</xref>], which reduces morbidity to the patient, possibility of installation through intraoral accesses, a lower rate of infections, possibility of a return to normal activities in a shorter period, satisfactory upper airway permeability and, mainly, reduction or absence of the need for intermaxillary block [<xref ref-type="bibr" rid="scirp.103879-ref3">3</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref5">5</xref>].</p><p>Proper stabilization of the fragments is important since bone repair is influenced by the dynamic forces of mastication [<xref ref-type="bibr" rid="scirp.103879-ref6">6</xref>]. Thus, the composition, manipu- lation, and adequate adaptation of the plates and screws are factors to be consi- dered in the stabilization of the bone segments, because of the dynamic comple- xity of the stomatognathic system [<xref ref-type="bibr" rid="scirp.103879-ref5">5</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref7">7</xref>].</p><p>Metals such as steel, vanadium, and titanium have been used in the manufacture of fastening materials. Titanium is used both pure and associated with other metals, and its modulus of elasticity allows adequate manipulation and adaptation of plates on the bone surface [<xref ref-type="bibr" rid="scirp.103879-ref8">8</xref>]. Despite this, excessive manipulation of the plates is expected to cause fatigue (a decrease in the strength limit of a metal, which may occur after repeated or cyclic loading, leading to metal fracture, with forces less than its original strength ) of the material, with its consequent rupture, being able to alter the stability of the fixation system [<xref ref-type="bibr" rid="scirp.103879-ref9">9</xref>]. Currently, with the variety of trademarks and composition of the fixation systems, it is necessary to evaluate the performance of these materials. The objective of the present study was to perform a comparative microscopic analysis of the resistance and hardness of internal fixation plates.</p></sec><sec id="s2"><title>2. Material and Methods</title><p>The research aimed to evaluate the flexural strength and hardness of the internal fixation materials through mechanical analysis by scanning electron microscopy.</p><p>The study sample consisted of 30 (thirty) 2.0 mm system plates, with four holes and bridge (intermediate portion without holes), divided into 3 groups, according to the manufacturer: Group 1: 10 plates of the Neoortho&#174; brand (Curitiba, PR, Brazil); Group 2: 10 Synthes&#174; (West Chester, PA, USA) brand name plates; Group 3: 10 Engimplan&#174; (Rio Claro, SP, Brazil) brand plates. It should be noted that the plates of Group 1 were donated by a distributor of the brand in Brazil, while the plates of Group 2 were donated by the Brazilian distributor and the plates of Group 3 were purchased from a distributor company accredited by the manufacturer, through the use of resources from the S&#227;o Paulo State Research Support Foundation (FAPESP) (Process number: 2011/062800), for another research project and also used for the present study.</p><p>In all plates of the 3 groups, the microscopic dimensions of the hardness and mechanical test were analyzed, as described below:</p><sec id="s2_1"><title>2.1. Dimensional Analysis</title><p>The purpose of the present analysis was to verify the standardization of the dimensions of the plates of the same manufacturer and between the manufacturers. The measurements were performed by a single examiner, according to a method modified by Brasileiro and Passeri [<xref ref-type="bibr" rid="scirp.103879-ref10">10</xref>], using a 150 mm/6&quot; digital caliper (<xref ref-type="fig" rid="fig1">Figure 1</xref>).</p><p>The measurements of each group were submitted to analysis of variation (ANOVA) and Turkey’s test with the significance of p &lt; 0.05. Data were analyzed using the Statistical software version 7 (State Soft Inc. Tulsa, USA).</p></sec><sec id="s2_2"><title>2.2. Mechanical Testing</title><p>The mechanical test was performed in the EMIC&#174; universal testing machine, model DL 500, which operates in a force range between 20 and 5000 N, with a three-point bending test method. The plates were positioned on the testing machine on two metal rollers and the force was applied by a load cell with a point incident on the center of the plate bridge. It should be noted that the testing machine performs the bending test to check the resistance by applying force and displacement of the measuring tip of the machine force. In the present study, only the plate was folded after the first application of the displacement; thus, after the first mechanical strength test, the plates were rectified and resubmitted to the same mechanical test, as described previously. The data obtained using the</p><p>mechanical test were submitted to analysis of variation (ANOVA) and the Tukey’s test with the significance of 95% (p &lt; 0.05), according to the studies of Trivellato, Mazzonetto, Passeri and Consani [<xref ref-type="bibr" rid="scirp.103879-ref11">11</xref>], and Brasileiro, Grempel, Ambrosano and Passeri [<xref ref-type="bibr" rid="scirp.103879-ref12">12</xref>]. Data were analyzed using the Statistical software version 7 (State Soft Inc. Tulsa, USA).</p></sec><sec id="s2_3"><title>2.3. Hardness Analysis (Vickers Hardness Test)</title><p>The plates were fixed (with Godiva) on a circular base of acrylic resin; after placing the test piece in the device (MicroMet 6040), the choice of placement was made to make the measurements through objective lenses with 10 and 50 x magnification. The apparatus was calibrated with a load of 500N and time of 10 seconds and the Vickers tip was selected. The computer program captures the image of the location where the indenter was printed (diamond-shaped) of the penetrator in the specimen, making it possible to measure the diagonals of the polygon to obtain the values of the test (HV) (<xref ref-type="fig" rid="fig2">Figure 2</xref>). The place of choice for application of the test was the corresponding area between holes at one end of the plate. Each test specimen was submitted to the hardness test at three different points, with a distance of 5 mm between them, where the established value was obtained through the average of these 3 points.</p></sec><sec id="s2_4"><title>2.4. Microscopic Analysis</title><p>The plates were subjected to scanning electron microscopy (Scanning Electron Microscope JSM-T220A, JEOL Ltd, Tokyo, Japan), in a 1000-fold increase in order to verify the components thereof, by reading the EDS (Energy-dispersive X-ray spectroscopy analysis) as well as for analysis of its surface structure.</p></sec></sec><sec id="s3"><title>3. Results</title><sec id="s3_1"><title>3.1. Dimensional Analysis</title><p>The analysis of dimensional standardization showed a statistically significant difference for all groups analyzed. The data related to the dimensional analysis are shown in <xref ref-type="table" rid="table1">Table 1</xref>.</p></sec><sec id="s3_2"><title>3.2. Mechanical Testing</title><p>The results indicate that there was a statistically significant difference (p &lt; 0.05) between the three groups when analyzed by the ANOVA test and that there were</p><table-wrap id="table1" ><label><xref ref-type="table" rid="table1">Table 1</xref></label><caption><title> Means and their respective standard deviations of the measurements of the plates. Analysis of variance between groups and the significance of ANOVA and Tukey’s tests</title></caption><table><tbody><thead><tr><th align="center" valign="middle"  rowspan="2"  >Medidas</th><th align="center" valign="middle"  colspan="3"  >M&#201;DIAS</th><th align="center" valign="middle"  rowspan="2"  >P &lt; 0.05</th></tr></thead><tr><td align="center" valign="middle" >Grupo 1</td><td align="center" valign="middle" >Grupo 2</td><td align="center" valign="middle" >Grupo 3</td></tr><tr><td align="center" valign="middle" >CT</td><td align="center" valign="middle" >29.741</td><td align="center" valign="middle" >27.91</td><td align="center" valign="middle" >30.488</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LF1D</td><td align="center" valign="middle" >4.454</td><td align="center" valign="middle" >4.98</td><td align="center" valign="middle" >5.344</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LF2D</td><td align="center" valign="middle" >4.426</td><td align="center" valign="middle" >4.98</td><td align="center" valign="middle" >5.337</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LF1E</td><td align="center" valign="middle" >4.434</td><td align="center" valign="middle" >4.97</td><td align="center" valign="middle" >5.368</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LF2E</td><td align="center" valign="middle" >4.407</td><td align="center" valign="middle" >4.97</td><td align="center" valign="middle" >5.364</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LFD</td><td align="center" valign="middle" >1.868</td><td align="center" valign="middle" >2.03</td><td align="center" valign="middle" >2.814</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LFE</td><td align="center" valign="middle" >1.843</td><td align="center" valign="middle" >2.02</td><td align="center" valign="middle" >2.804</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LPFD</td><td align="center" valign="middle" >2.367</td><td align="center" valign="middle" >2.5</td><td align="center" valign="middle" >2.819</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LP</td><td align="center" valign="middle" >2.312</td><td align="center" valign="middle" >2.5</td><td align="center" valign="middle" >2.811</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >LPFE</td><td align="center" valign="middle" >2.308</td><td align="center" valign="middle" >2.49</td><td align="center" valign="middle" >2.817</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >EFD</td><td align="center" valign="middle" >2.116</td><td align="center" valign="middle" >3.94</td><td align="center" valign="middle" >4.239</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >EFE</td><td align="center" valign="middle" >2.063</td><td align="center" valign="middle" >3.95</td><td align="center" valign="middle" >4.214</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >CP</td><td align="center" valign="middle" >8.209</td><td align="center" valign="middle" >8.94</td><td align="center" valign="middle" >10.023</td><td align="center" valign="middle" >All groups</td></tr><tr><td align="center" valign="middle" >DF1D</td><td align="center" valign="middle" >1.994</td><td align="center" valign="middle" >2.01</td><td align="center" valign="middle" >2.079</td><td align="center" valign="middle" >G1 e G3/G2 e G3</td></tr><tr><td align="center" valign="middle" >DF2D</td><td align="center" valign="middle" >2.011</td><td align="center" valign="middle" >2.01</td><td align="center" valign="middle" >2.081</td><td align="center" valign="middle" >G1 e G3/G2 e G3</td></tr><tr><td align="center" valign="middle" >DF1E</td><td align="center" valign="middle" >2.006</td><td align="center" valign="middle" >2.02</td><td align="center" valign="middle" >2.089</td><td align="center" valign="middle" >G1 e G3/G2 e G3</td></tr><tr><td align="center" valign="middle" >DF2E</td><td align="center" valign="middle" >2.006</td><td align="center" valign="middle" >2.02</td><td align="center" valign="middle" >2.071</td><td align="center" valign="middle" >G1 e G3/G2 e G3</td></tr><tr><td align="center" valign="middle" >EspF1D</td><td align="center" valign="middle" >0.907</td><td align="center" valign="middle" >1.01</td><td align="center" valign="middle" >0.983</td><td align="center" valign="middle" >G1 e G2/G1 e G3</td></tr><tr><td align="center" valign="middle" >EspF2D</td><td align="center" valign="middle" >0.923</td><td align="center" valign="middle" >1</td><td align="center" valign="middle" >0.957</td><td align="center" valign="middle" >G1 e G2</td></tr><tr><td align="center" valign="middle" >EspFD</td><td align="center" valign="middle" >0.919</td><td align="center" valign="middle" >1</td><td align="center" valign="middle" >0.952</td><td align="center" valign="middle" >G1 e G2</td></tr><tr><td align="center" valign="middle" >EspPD</td><td align="center" valign="middle" >0.923</td><td align="center" valign="middle" >1</td><td align="center" valign="middle" >0.966</td><td align="center" valign="middle" >N&#227;o Significante (P &gt; 0.05)</td></tr><tr><td align="center" valign="middle" >EspCP</td><td align="center" valign="middle" >0.875</td><td align="center" valign="middle" >1</td><td align="center" valign="middle" >0.97</td><td align="center" valign="middle" >G1 e G2/G1 e G3</td></tr><tr><td align="center" valign="middle" >EspPE</td><td align="center" valign="middle" >0.922</td><td align="center" valign="middle" >1</td><td align="center" valign="middle" >0.969</td><td align="center" valign="middle" >G2 e G3</td></tr><tr><td align="center" valign="middle" >EspF1E</td><td align="center" valign="middle" >0.932</td><td align="center" valign="middle" >1</td><td align="center" valign="middle" >0.956</td><td align="center" valign="middle" >G1 e G3</td></tr><tr><td align="center" valign="middle" >EspF2E</td><td align="center" valign="middle" >0.923</td><td align="center" valign="middle" >1</td><td align="center" valign="middle" >0.957</td><td align="center" valign="middle" >G1 e G3</td></tr><tr><td align="center" valign="middle" >EspFE</td><td align="center" valign="middle" >0.864</td><td align="center" valign="middle" >1</td><td align="center" valign="middle" >0.959</td><td align="center" valign="middle" >G2 e G3</td></tr></tbody></table></table-wrap><p>Legend: CT—Total length of the plate, LF1D—Width of the holes 1D, LF2D—Width of the holes 2D, LF1E—Width of the holes 1E, LF2E—Width of the holes 2E, LFD-LFE—Width region between holes E, LPFD—Width intermediate region/bridge hole D, LP—Width intermediate region/center bridge, LPFE— Width intermediate region/bridge hole E, EFD—Space between holes D, EFE—Space between holes E, CP—Bridge length, DF1D—Diameter of hole 1D, DF2D—Diameter of hole 2D, DF1E—Diameter of hole 1E, DF2E—Diameter of hole 2E, EspF1D—Plate thickness over hole 1D, EspF2D—Plate thickness on thickness 2D, EspFD—Thickness of the plate between holes D, EspPD—Thickness of the bridge plate at D, EspCP—Thickness of the bridge center plate, EspPE—Thickness of the bridge plate at E, EspF1E—Thickness of the plate on hole 1E, EspF2E—Plate thickness over hole 2E, EspFE—Plate thickness between holes E.</p><p>intra-group differences between the first and second stages between Groups 2 and 3, that is, Group 2 presented higher mechanical resistance when compared to Groups 1 and 3. <xref ref-type="table" rid="table2">Table 2</xref> shows the mean values obtained in the mechanical test, as well as the results of the statistical tests used (ANOVA and Tukey’s p &lt; 0.05), while Graph 1 shows the relationship of resistance comparing between the three groups.</p></sec><sec id="s3_3"><title>3.3. Vickers Hardness Analysis</title><p>In general, it was observed that the plates of Group 2 were harder than those of the other groups. <xref ref-type="table" rid="table3">Table 3</xref> shows the values for the total mean of each group.</p></sec><sec id="s3_4"><title>3.4. Analysis of Electronic Microscopy and EDS</title><p>The analysis of the electron microscopy showed that the three groups present different aspects of surface finishing when viewed with 1000x increase (<xref ref-type="fig" rid="fig3">Figure 3</xref>).</p><p>EDS showed that the plates of the three groups studied had a similar composition, being composed basically of commercially pure titanium (Ti 98%) Grade 1. The plates of Group 3 contained phosphorus particles in their composition and the aluminum detected in the reading took place of the support that sustained the samples and were not relevant to the study. The analyses are represented in Graph 2, Graph 3, and Graph 4.</p><table-wrap id="table2" ><label><xref ref-type="table" rid="table2">Table 2</xref></label><caption><title> Mean values of the mechanical test, in Newtons, standard deviation (S), difference between F2 and F1 and test results of intra and intergroup variance (ANOVA) and Tukey’s test (p &lt; 0.05)</title></caption><table><tbody><thead><tr><th align="center" valign="middle" ></th><th align="center" valign="middle" >F1 (S)</th><th align="center" valign="middle" >F2 (S)</th><th align="center" valign="middle" >F2-F1</th></tr></thead><tr><td align="center" valign="middle" >Group 1</td><td align="center" valign="middle" >16.05 (0.41)</td><td align="center" valign="middle" >16.42 (0.47)</td><td align="center" valign="middle" >0.37*</td></tr><tr><td align="center" valign="middle" >Group 2</td><td align="center" valign="middle" >23.6 (0.33)</td><td align="center" valign="middle" >24.92 (0.31)</td><td align="center" valign="middle" >1.32</td></tr><tr><td align="center" valign="middle" >Group 3</td><td align="center" valign="middle" >17.03 (0.34)</td><td align="center" valign="middle" >17.9 (0.30)</td><td align="center" valign="middle" >0.87</td></tr></tbody></table></table-wrap><p>Legend: F1 = force in the first measurement; F2 = force in the second measurement. * = not significant in the intragroup evaluation.</p><disp-formula id="scirp.103879-formula1"><graphic  xlink:href="//html.scirp.org/file/4-1460854x4.png"  xlink:type="simple"/></disp-formula><p>Graph 1. Measurement of the loads obtained during the flexion test. F1 = first test; F2 = second test after manual rectification of the plate.</p><table-wrap id="table3" ><label><xref ref-type="table" rid="table3">Table 3</xref></label><caption><title> Mean and standard deviation (SD) of Vickers hardness (HV) and statistical significance. S = significant (p &lt; 0.05)</title></caption><table><tbody><thead><tr><th align="center" valign="middle" >Grupos</th><th align="center" valign="middle" >M&#233;dia (DP)</th><th align="center" valign="middle" >Signific&#226;ncia</th></tr></thead><tr><td align="center" valign="middle" >1</td><td align="center" valign="middle" >199.39 (3.32)</td><td align="center" valign="middle" >S</td></tr><tr><td align="center" valign="middle" >2</td><td align="center" valign="middle" >251.31 (2.88)</td><td align="center" valign="middle" >S</td></tr><tr><td align="center" valign="middle" >3</td><td align="center" valign="middle" >202.62 (2.66)</td><td align="center" valign="middle" >S</td></tr></tbody></table></table-wrap><disp-formula id="scirp.103879-formula2"><graphic  xlink:href="//html.scirp.org/file/4-1460854x6.png"  xlink:type="simple"/></disp-formula><p>Graph 2. Surface composition of Group 1 plates.</p><disp-formula id="scirp.103879-formula3"><graphic  xlink:href="//html.scirp.org/file/4-1460854x7.png"  xlink:type="simple"/></disp-formula><p>Graph 3. Surface composition of Group 2 plates.</p><disp-formula id="scirp.103879-formula4"><graphic  xlink:href="//html.scirp.org/file/4-1460854x8.png"  xlink:type="simple"/></disp-formula><p>Graph 4. Surface composition of Group 3 plates.</p></sec></sec><sec id="s4"><title>4. Discussion</title><p>It is known that for the repair of bone fractures and osteotomies, adequate immobilization of the fragments is essential. In Buccomaxillofacial Surgery and Traumatology, the use of plate and screw systems has been the method of elective fixation for the treatment of face fractures and for the immobilization of bone fragments after osteotomies. There is a great diversity of fixation systems today; these, vary not only the dimensions of their plates and screws, but also their designs and constituents.</p><p>The present study was born from the impression, during use, that some plates were softer than others, since they seemed to more easily allow the folds necessary to adapt them to the bone tissue. Such adaptation may interfere with the mechanical resistance of the plates [<xref ref-type="bibr" rid="scirp.103879-ref13">13</xref>]. Therefore, one of the objectives was to evaluate the mechanical strength of 3 plate systems, 2.0 mm. Obviously, the dimensions of the fixation system employed influence both the execution of folds for adaptation and the immobilization of the bone. We chose to use 2.0 mm system plates, with intermediate (bridge) and 4 holes, which are plates widely employed in the fixation of mandibular osteotomies. It should be noted that, in this context, the standardization of the dimensions of the plates has to be verified.</p><p>Therefore, before carrying out the study of the resistance of the plates, the dimensional analysis of the plates was carried out to verify if plates of different manufacturers, all of the 2.0 mm system, had the same dimensions, since dimensional differences of the systems can influence the mechanical test [<xref ref-type="bibr" rid="scirp.103879-ref9">9</xref>]. The results of the present study are in agreement with other studies [<xref ref-type="bibr" rid="scirp.103879-ref9">9</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref11">11</xref>], which verified that there are dimensional differences between the plates of different manufacturers and that this seems to be more critical for the plates manufactured in Brazil (Groups 1 and 2 of the present study). Although statistically significant, the observed differences were small between the plates of Groups 1 and 2; the plates of Group 3 presented greater dimensional difference.</p><p>Despite the above, in the present study, this difference did not seem to have influenced the mechanical resistance, since the analysis of flexion of the plates of Group 3 were less resistant than those of Group 2, and more resistant than that of Group 1. It is worth noting that the plates of Group 2 showed lower variability among their dimensions, besides being smaller than those of Group 3, and presented the highest mechanical resistance among the 3 groups studied. The results of the flexion test (mechanical resistance test of the present study, according to the study by Rocha J&#250;nior et al. [<xref ref-type="bibr" rid="scirp.103879-ref9">9</xref>], who also tested plates of the same brands of the plates of Groups 1 and 2, and with Trivellato et al. [<xref ref-type="bibr" rid="scirp.103879-ref11">11</xref>] who observed higher values (in the present study), when compared to the values of the Group 3 plate flexion test (from the present study). It should be emphasized that the study by Trivellato et al. [<xref ref-type="bibr" rid="scirp.103879-ref11">11</xref>] evaluated, among others, the same plates of Groups 2 and 3 of the present study.</p><p>The hardness of a material can be translated by the penetration resistance of its surface. The Vickers hardness is based on the resistance the material offers to the penetration of a square-shaped diamond pyramid and the angle between faces of 136˚ under a given load. The Vickers hardness value (HV) is the quotient of the applied load (F) by the printing area (A) left on the body tested. The study by Matthew et al. [<xref ref-type="bibr" rid="scirp.103879-ref14">14</xref>] compared the Vickers hardness of titanium plates and stainless steel, which validates the method for Vickers plate hardness verification; besides, their results showed Vickers hardness difference between the studied plates, which corroborates the importance of the components of the plate about its hardness. It seems logical to believe that harder plates, that is, with greater Vickers hardness, are more resistant and possibly result in a more stable fixation. Also, because they are harder, they can become more difficult to bend and adapt to the bone. In the present study, the plates of Group 2 showed Vickers hardness superior to the plates of Groups 1 and 3, this being a factor that contributed for the increased resistance of the plates of this group, even being dimensionally smaller, when compared with those of Group 3, for example. This fact may be due to its composition, even though it is more similar to the composition of the plates of Group 1. If the plates are of similar composition, perhaps the manufacturing process is responsible for the hardness difference. We lack data to deepen the discussion in this regard.</p><p>The values obtained in the evaluation of mechanical strength can be attributed to the design, standardization of dimensions, and constituent material. Thus, the higher mechanical resistance of plates of different sizes may be due to differences between the constituent components of the plates [<xref ref-type="bibr" rid="scirp.103879-ref15">15</xref>] [<xref ref-type="bibr" rid="scirp.103879-ref16">16</xref>]. The energy dispersive spectrometry (EDS) performed in the present study qualitatively showed the chemical elements of the surface of the plates. The present study is in agreement with Trivellato et al. [<xref ref-type="bibr" rid="scirp.103879-ref11">11</xref>], because, in a similar way to the authors mentioned above, the presence of commercially pure titanium was verified, classified as grade 1, in all plates evaluated, however, phosphorus and oxygen ions were found in Group 3; it is worth noting that the study by Trivellato et al. [<xref ref-type="bibr" rid="scirp.103879-ref11">11</xref>] evaluated plates of markings corresponding to Groups 2 and 3 of the present study. On the other hand, Rocha J&#250;nior et al. [<xref ref-type="bibr" rid="scirp.103879-ref9">9</xref>], evidenced the presence of pure titanium in the samples of the same manufacturers of Groups 1 and 2 of the present study, plus sodium, phosphorus, sulfur, potassium, chlorine, carbon, nitrogen, aluminum, silicon and calcium, iron, zinc, phosphorus, potassium, nickel, and nitrogen (on the plates of Group 2 of the present study).</p><p>The presence of aluminum observed in the SDS of Groups 1 and 2 of the present study may be due to external contamination, specifically the tip of the universal testing machine. Thus, since the constituent components of the three groups are basically the same, the discrete differences found do not seem to be responsible for the differences in the results of the mechanical tests. Perhaps the manufacturing process may have some interference in this, but we lack data to discuss this process, since the manufacturers usually keep secrecy of their pro- cesses or part of them.</p></sec><sec id="s5"><title>5. Conclusion</title><p>It was concluded that, although the studied systems presented similar compounds, it was possible to observe different behaviors in terms of size, strength, hardness, and surface characteristics. The plates of Group 2 presented a superior performance about the dimensional analysis and the tests of strength and hardness, indicating that it can present greater stability to the bone fragments fixed with it. We also consider that the clinical applicability of the results varies according to the level of professional experience, relating in large part according to the number of movements performed during the manipulation of the material. Thus, it is observed that no material has excellent characteristics in all necessary aspects, and it is necessary to evaluate the favorable points that bring the expected benefits to the needs of the patient and the surgeon.</p></sec><sec id="s6"><title>Conflicts of Interest</title><p>The authors declare no conflicts of interest regarding the publication of this paper.</p></sec><sec id="s7"><title>Cite this paper</title><p>Neto, V.T., Fia- moncini, E.S., Ferreira, M.F., Flumignan, D.L., Vaz, L.G. and Gon&#231;ales, E.S. (2020) Comparative Dimensional and Microscopic Analysis of the Strength and Hardness of Internal Fixation Plates. Open Journal of Stomatology, 10, 310-320. https://doi.org/10.4236/ojst.2020.1010029</p></sec></body><back><ref-list><title>References</title><ref id="scirp.103879-ref1"><label>1</label><mixed-citation publication-type="other" xlink:type="simple">Michelet, F.X., Deymes, J. and Dessus, B. (1973) Osteosynthesis with Miniaturized Screwed Plates in Maxillo-Facial Surgery. Journal of Maxillofacial Surgery, 1, 79-84. https://doi.org/10.1016/S0301-0503(73)80017-7</mixed-citation></ref><ref id="scirp.103879-ref2"><label>2</label><mixed-citation publication-type="other" xlink:type="simple">Ellis, E. (1993) Rigid Skeletal Fixation of Fractures. Journal of Oral and Maxillofacial Surgery, 51, 163-173. https://doi.org/10.1016/S0278-2391(10)80016-3</mixed-citation></ref><ref id="scirp.103879-ref3"><label>3</label><mixed-citation publication-type="other" xlink:type="simple">Obwegeser, H.L. (2007) Orthognathic Surgery and a Tale of How Three Procedures Came to Be: A Letter to the Next Generations of Surgeons. Clinics in Plastic Surgery, 34, 331-355. https://doi.org/10.1016/j.cps.2007.05.014</mixed-citation></ref><ref id="scirp.103879-ref4"><label>4</label><mixed-citation publication-type="other" xlink:type="simple">Ellis III, E. (2014) An Algorithm for the Treatment of Noncondylar Mandibular Fractures. Journal of Oral and Maxillofacial Surgery, 72, 939-949. https://doi.org/10.1016/j.joms.2013.11.026</mixed-citation></ref><ref id="scirp.103879-ref5"><label>5</label><mixed-citation publication-type="other" xlink:type="simple">Al-Moraissi, E.A. and Ellis, E. (2014) What Method for Management of Unilateral Mandibular Angle Fractures Has the Lowest Rate of Postoperative Complications? A systematic Review and Meta-Analysis. Journal of Oral and Maxillofacial Surgery, 72, 2197-2211. https://doi.org/10.1016/j.joms.2014.05.023</mixed-citation></ref><ref id="scirp.103879-ref6"><label>6</label><mixed-citation publication-type="other" xlink:type="simple">Feller, K.U.G., Richter, M. and Schneider, U. (2002) Combination of Microplate and Miniplate for Osteosynthesis of Mandibular Fractures: An Experimental Study. International Journal of Oral and Maxillofacial Surgery, 31, 78-83.https://doi.org/10.1054/ijom.2000.0182</mixed-citation></ref><ref id="scirp.103879-ref7"><label>7</label><mixed-citation publication-type="other" xlink:type="simple">Ellis III, E. and Esmail, N. (2009) Malocclusions Resulting from Loss of Fixation after Sagittal Split Ramus Osteotomies. Journal of Oral and Maxillofacial Surgery, 67, 2528-2533. https://doi.org/10.1016/j.joms.2009.06.022</mixed-citation></ref><ref id="scirp.103879-ref8"><label>8</label><mixed-citation publication-type="other" xlink:type="simple">Roig, M.A. (2002) Valoración a largo plazo de los resultados de tratamiento mediante osteosíntesis con miniplacas en sus diferentes usos en cirugía craneomaxilofacial. Tese (Doutorado em Cirurgia e Medicina)—Facultat de Medicina, Barcelona, 138 p.</mixed-citation></ref><ref id="scirp.103879-ref9"><label>9</label><mixed-citation publication-type="other" xlink:type="simple">Rocha Júnior, H.V., Padovan, L.E.M., Oliveira, R.B., Assis, D.S.F.R. and Duarte, M.A.H. (2013) Avalia&amp;#231;&amp;#227;o ultraestrutural,dimendional e mecanica de dois sistemas de fixa&amp;#231;&amp;#227;o interna rígida de 2mm. Revista Salusvita, 32, 125-137.</mixed-citation></ref><ref id="scirp.103879-ref10"><label>10</label><mixed-citation publication-type="other" xlink:type="simple">Brasileiro, B.F. and Passeri, L.A. (2008) Avalia&amp;#231;&amp;#227;o dimensional de miniplacas e parafusos para cirurgia ortognática. Revista Brasileira de Cirurgia e Traumatologia, 5, 51-56.</mixed-citation></ref><ref id="scirp.103879-ref11"><label>11</label><mixed-citation publication-type="other" xlink:type="simple">Trivellato, A.E., Mazzonetto, R., Passeri, L.A. and Consani, S. (2000) Estudo químico, macroscópico e da resistência à flex&amp;#227;o de placas e parafusos de titanio usados na fixa&amp;#231;&amp;#227;o interna rígida. Pesquisa Odontológica Brasileira, 14, 392-398.https://doi.org/10.1590/S1517-74912000000400015</mixed-citation></ref><ref id="scirp.103879-ref12"><label>12</label><mixed-citation publication-type="other" xlink:type="simple">Brasileiro, B.F., Grempel, R.G., Ambrosano, G.M.B. and Passeri, L.A. (2009) An in Vitro Evaluation of Rigid Internal Fixation Techniques for Sagittal Split Ramus Osteotomies: Advancement Surgery. Journal of Oral and Maxillofacial Surgery, 67, 809-817. https://doi.org/10.1016/j.joms.2008.11.009</mixed-citation></ref><ref id="scirp.103879-ref13"><label>13</label><mixed-citation publication-type="other" xlink:type="simple">Haug, R.H., Street, C.C. and Goltz, M. (2002) Does Plate Adaptation Affect Stability? A Biomechanical Comparison of Locking and Nonlocking Plates. Journal of Oral and Maxillofacial Surgery, 60, 1319-1326.https://doi.org/10.1053/joms.2002.35732</mixed-citation></ref><ref id="scirp.103879-ref14"><label>14</label><mixed-citation publication-type="other" xlink:type="simple">Matthew, I.R., Frame, J.W., Browne, R.M. and Millar, B.G. (1996) In Vivo Surface Analysis of Titanium and Stainless Steel Miniplates and Screws. International Journal of Oral and Maxillofacial Surgery, 25, 463-468.https://doi.org/10.1016/S0901-5027(96)80085-3</mixed-citation></ref><ref id="scirp.103879-ref15"><label>15</label><mixed-citation publication-type="other" xlink:type="simple">De Oliveira, K.P., De Moraes, P.H., Da Silva, J.S.P., De Queiroz, W.F. and Germano, A.R. (2014) In Vitro Mechanical Assessment of 2.0-mm System Three-Dimensional Miniplates in Anterior Mandibular Fractures. International Journal of Oral and Maxillofacial Surgery, 43, 564-571. https://doi.org/10.1016/j.ijom.2013.10.008</mixed-citation></ref><ref id="scirp.103879-ref16"><label>16</label><mixed-citation publication-type="other" xlink:type="simple">Kun&amp;#269;ická, K., Kocich, R. and Lowe, T.C. (2017) Advances in Metals and Alloys for Joint Replacement. Progress in Materials Science, 88, 232-280. https://doi.org/10.1016/j.pmatsci.2017.04.002</mixed-citation></ref></ref-list></back></article>