Smilowitz, J.B., Das, I.J., Feygelman, V., Fraass, B.A., Geurts, M., Kry, S.F., et al. (2016) AAPM Medical Physics Practice Guideline 5.a.: Commissioning and QA of Treatment Planning Dose Calculations—Megavoltage Photon and Electron Beams. Journal of Applied Clinical Medical Physics, 17, 457.
https://doi.org/10.1120/jacmp.v17i1.6166
has been cited by the following article:
-
TITLE:
TrueBeam Low Dose Rate Investigation for Pulsed Reduced Dose Rate IMRT
AUTHORS:
Mark Geurts
KEYWORDS:
PRDR, Low Dose Rate, Delta 4, TrueBeam
JOURNAL NAME:
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology,
Vol.6 No.2,
May
22,
2017
ABSTRACT: The
capability of the TrueBeam treatment system to deliver step and shoot IMRT
plans at low dose rates was evaluated. Beam characteristics during low dose
rates (5 to 100 MU/min) were evaluated for consistency using a planar ion
chamber array. MU constancy, linearity, and beam profile symmetry were all
found to be equivalent within 0.5%. The response of the Scandi Dos Delta 4
system was also evaluated at low dose rates of using static open beams compared
to ion chamber measurements, and step and shoot IMRT plans comparing 5 - 20
MU/min and 100 MU/min dose rates, with a maximum observed absolute dose
difference of 0.8% and equivalence margin of 0.2%. The Gamma Index and
measurement reproducibility were also found to be equivalent.