Article citationsMore>>
Kang, S., Lang, J., Wang, P., Li, J., Lin, M., Chen, X., et al. (2014) Optimization Strategies for Pulsed Low-Dose-Rate IMRT of Recurrent Lung and Head and Neck Cancers. Journal of Applied Clinical Medical Physics, 15, 4661.
https://doi.org/10.1120/jacmp.v15i3.4661
has been cited by the following article:
-
TITLE:
TrueBeam Low Dose Rate Investigation for Pulsed Reduced Dose Rate IMRT
AUTHORS:
Mark Geurts
KEYWORDS:
PRDR, Low Dose Rate, Delta 4, TrueBeam
JOURNAL NAME:
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology,
Vol.6 No.2,
May
22,
2017
ABSTRACT: The
capability of the TrueBeam treatment system to deliver step and shoot IMRT
plans at low dose rates was evaluated. Beam characteristics during low dose
rates (5 to 100 MU/min) were evaluated for consistency using a planar ion
chamber array. MU constancy, linearity, and beam profile symmetry were all
found to be equivalent within 0.5%. The response of the Scandi Dos Delta 4
system was also evaluated at low dose rates of using static open beams compared
to ion chamber measurements, and step and shoot IMRT plans comparing 5 - 20
MU/min and 100 MU/min dose rates, with a maximum observed absolute dose
difference of 0.8% and equivalence margin of 0.2%. The Gamma Index and
measurement reproducibility were also found to be equivalent.