![]() Open Journal of Nursing, 2012, 2, 307-310 OJN http://dx.doi.org/10.4236/ojn.2012.223045 Published Online November 2012 (http://www.SciRP.org/journal/ojn/) Comparison of temperature measurements in bladder, rectum and pulmonary artery in patients after cardiac surgery Hendrika Wollerich, Farouq Ismael, Maarten W. Nijsten, Willem Dieperink Department of Critical Care, University of Groningen, University Medical Center Groningen, Groningen, Netherlands Email: [email protected] Received 10 September 2012; revised 12 October 2012; accepted 25 October 2012 ABSTRACT In many patients in the intensive care unit (ICU) con- tinuous temperature monitoring is performed with rectal probes. Currently there are more options to measure temperature in critically ill patients. Since bladder catheters are routinely used in the majority of ICU patients, using bladder temperature (Tb) could do away with rectal probes. In this prospective study, we compared Tb, rectal temperature (Tr) and pulmonary artery catheter temperature (Tpa) in pa- tients who underwent cardiac surgery. We also eva lu- ated if urinary production affected Tb. Patients ad- mitted af ter cardiac surg ery with sensors in place for measurement of Tr, Tb and Tpa upon arrival at the ICU were included. Diuresis was recorded every hour. Data were collected until 24 h after admission or until ICU discharge. Nineteen consecutive patients (9 males; mean age 61 years) were evaluated. Over a median observation period of 17 hours 382 Tb, 333 Tr and 371 Tpa measurements were recorded. Linear corre- lations (R) between Tb and Tr, between Tb and Tpa and between Tr and Tpa were 0.95, 0.95 and 0.91 re- spectively (P < 0.001). Bland-Altman analysis demon- strated no relation between temperature and the (Tb- Tr) offset. No relation of diuresis with (Tb-Tr) or (Tb-Tpa) was observed. After cardiac surgery, blad- der temperature performed as well as conventional rectal probes with no interference of diuresis on blad- der temperature measurement. Thus, the use of blad- der temperature probes may be preferable to rectal probes in patients after cardiac surgery. Keywords: Intensive Care; Cardiac Surgery; Body Temperature 1. INTRODUCTION Accurate and precise measurement of body temperature is important for in the timely detection of fever or hypo- thermia in patients in the intensive care unit (ICU). No uniform practice regarding temperature measurement exists. A variety of devices are used to measure body temperature at diverse anatomical locations in two possi- ble temperature scales. Traditionally the temperature meas- ured within the pulmonary artery (Tpa) with a the Swan- Ganz catheter (PAC), that is typically only used in pa- tients with complex hemodynamic problems, is consid- ered the “gold standard” for the measurement of body temperature [1]. Nevertheless, for routine clinical use in ICU patients, using a PAC is clearly inapplicable. In many ICU’s continuous rectal temperature meas- urement (Tr) is standard practice [2]. Placement of ther- mometers in the rectum is more invasive than in the axilla or groin. A rectal probe can be distressing as well as embarrassing for patients. Moreover, for inserting the probe, patients must be placed on their side, a maneuver that often requires two nurses to safely turn the patient. The rectal temperature probe can also result in mucosal injury, or even perforation of the rectum [3]. Tr can also be affected by the passage of stools and the probe can dislocate easily. On the other hand Tr is more accurate than less invasive measurement methods such as at the axilla or inguinal [4]. Urinary bladder temperature (Tb), measurement of temperature with a specially designed temperature probe embedded in a Foley urinary catheter, could be a favor- able alternative to Tr in the ICU because rectal probe insertion is not required without the need for additional procedures [5,6]. In several patient groups it has been established that Tb generally corresponds well to the Tpa, the “gold standard” [7,8]. It has been suggested, that the accuracy and precision of Tb might be influenced by the urinary flow rate [5,9]. In 22 ICU patients in which tem- perature was simultaneously monitored with PAC it has been demonstrated that the reliability of Tb is higher than either Tr or inguinal or axillary temperature measure- ments [6,10]. OPEN ACCESS ![]() H. Wollerich et al. / Open Journal of Nursing 2 (2012) 307-310 308 It is not known if the same holds true for the reliability of Tb during the direct post-operative phase in patients admitted after cardiac surgery. Cardiac surgery involves the opening of the chest and frequently the use of a heart-lung machine. Thus during and after such surgery important swings in hemodynamics and temperature can occur. This may be particularly relevant given the large thermal gradients from core to periphery that occur dur- ing anesthesia and cardiac surgery [5]. Objective In a prospective study we tested the hypothesis that after cardiac surgery Tb adequately reflects Tpa, and at least as good as Tr. We also examined the effect of varying urinary flow rate on the Tb. 2. METHODS During a 10-month study period (June 2009 to March 2010) consecutive patients aged >18 years, who were undergoing cardiac surgery and admitted to the depart- ment of critical care were evaluated for this study. Pa- tients, with a ileal conduit urinary diversion also known as Bricker ileal conduit, or patients with suprapubic catheters, or other patients where urethral catheterization was contraindicated as well as patients who were under- going renal replacement therapy were excluded. Only patients with both a PAC, a urinary bladder catheter with temperature probe (Tb) and a rectal tem- perature probe (Tr) were included. Tb was measured by a probe incorporated in a Foley urinary catheter (DeRoyal, Powell, USA). Tpa was meas- ured by a pulmonary artery catheter (PAC) (Edwards, Irvine California). Tr was measured by a digital ther- mometer probe (Esophageal/Rectal temperature probe 25˚C - 45˚C, Philips Medical Systems, Andover, USA) designed for axillar and rectal use. The temperature sen- sors and the PAC were respectively connected to a tem- perature and cardiac output module of the bedside moni- tor (MP70, Philips Medical Systems, Eindhoven, the Netherlands). Temperature values were recorded in de- grees Celsius every 30 minutes by the ICU nurses. The urinary production in ml was recorded every hour. All data were collected until 24 h after admission or until ICU discharge if this took place within 24 h. Tpa was considered the reference value. Thus Tb was compared with Tpa, and Tr was compared with Tpa us- ing regression and Bland and Altman analysis. Since no additional interventions were performed in- formed consent was not required for this anonymized analysis of data. The institutional review board of the University Medical Center Groningen approved this ob- servational study (METc 2008/279). 3. RESULTS During the study period, 19 consecutive patients admit- ted to the department of critical care after cardiac surgery were included. For 9 men and 10 women with a mean age of 61 ± 13 (SD) years, we recorded temperatures over a median observation period of 17 hours. A total of 382 Tb, 333 Tr and 371 Tpa measurements were ana- lyzed (Table 1). We also recorded 291 diuresis meas- urements. Correlation coefficients between Tb and Tr, between Tb and Tpa (Figure 1) and between Tr and Tpa were 0.95, 0.95 and 0.91 respectively (P < 10−9). The mean offsets for expressed as (Tb-Tr), (Tb-Tpa) and (Tr-Tpa) were −0.05˚C, +0.05˚C and +0.10˚C respectively, indi- cating that Tr was the least reliable of the three meas- urements. Bland-Altman analysis demonstrated no relation be- tween temperature and the (Tb-Tr) offset (Figure 2). No relation of diuresis with (Tb-Tr) or (Tb-Tpa) was ob- served. The two outlying values visible in Figures 1 and 2 concerned immediate post-operative patients with very cold peripheries and consequently high temperature gra- dients. For both these outliers Tr deviated more from Tpa than Tb. 4. DISCUSSION Measuring body temperature in critically ill patients is a part of routine nursing care. However, it is still unclear which method and device is ideal for measuring accurate, reliable affordable, comfortable and safe body tempera- ture. Nurses therefore must have the knowledge and skills related to the various temperature measuring Table 1. Patient characteristics. Number of patients 19 Age, mean, SD, (range) 61.1 ± 13.4 (32 - 83) Gender, male/female 9/10 Admission diagnoses: Valve replacement, n (%) CABG, n (%) Cardiovascular other, n (%) 9 (47.3) 5 (26.3) 5 (30.3) Height (cm), mea n , SD, (rang e ) 171.1 ± 8.2 (156 - 189) Weight (kg), mean, SD, (range) 79.0 ± 16.2 (51 - 105) BSA (m2), mean, SD, (range) 1.9 ± 0.2 (1.53 - 2.37) APACHE IV score, me an, ±SD, (range) 44.8 ± 15.7 SAPS II score, mean, ±SD 30.5 ± 8.1 Diuresis, ml/hr, mean, ±SD, (range) 88 ± 79 (5 - 400) SD = Standard Deviation, CABG = Coronary Artery Bypass Graft Surgery, BSA = Body Surface Area, APACHE II = Acute Physiologic and Chronic Health Evaluation, SAPS II = Simplified Acute Physiology Score. Copyright © 2012 SciRes. OPEN ACCESS ![]() H. Wollerich et al. / Open Journal of Nursing 2 (2012) 307-310 309 393837363534 39 38 37 36 35 34 PA C temperature Bl a dder te mperatur e Co rre la tion of bl a dder t e mpe ra tur e versus P A C te mpe ra tur e R=0.95 (P<10-9) Figure 1. Correlation plot of Tb versus Tpa showing high cor- relation between the two temperature measuring methods. 393837363534 1 0 -1 -2 -3 MEAN OF BOTH METHO DS, degrees Celsius DIFFERENCES BETWEEN METHODS, degrees Celsius Blad der temperature versus PAC t emperature Figure 2. Bland and Altman Plot of the data obtained from paired samples of bladder and PAC temperature. Correlation R = 0.95 (P < 10−9). methods and associated devices to make good clinical decisions. In the ICU, Tb measurement has gained in- creased acceptance and is simplified by the fact that most patients have an indwelling Foley catheter. In many clinical studies Tpa is considered as the most optimal representation of internal body temperature [8, 11,12]. However, for temperature measurement only, Tpa is obviously not a practical technique as the invasive pulmonary artery catheter is only required in a minority of ICU patients. Inserting rectal thermometers is usually uncomplicated, although it can occasionally cause mu- cosal injury. However, these probes can be easily dislo- cated by stool movements or by the application of sup- positories and they are definitely associated with addi- tional patient discomfort. At our ICU most patients have bladder catheters, making the use of a catheter with an incorporated temperature probe therefore not additionally invasive. Thus Tb could be the preferred temperature device in ICU patients. The data in our study agree with those found in other patient groups where neither Tb nor Tr exactly correlate with Tpa [13-17]. However, bladder temperature gener- ally did as well or better compared to Tr. In spite of the fact that devices for Tb measurement are more expensive than Tr devices, the overall costs for Tb are lower because Tr probes must be changed more often than Tb probes. Thus, although overall costs may vary according to local circumstances, in our setting the use of Tb is associated with reduced yearly cost decrease of more than €4000 for our unit. There are some limitations of the present study. First, none of the CE-certified thermometers were tested or compared with certified laboratory thermometers. On the other hand, all equipment and materials in our hospital are routinely monitored and checked regularly, and Tpa is regarded as the gold standard. Second, rectal temperature measurement could be af- fected by the depth of inserting the probe, conditions affecting local blood flow and the presence of stools. These factors were not assessed in the our study since we assumed that interference of stools or probe dislocation would be less important, so early after insertion and sur- gery. But if these factors come into play later on, we ex- pect them to work particularly to the disadvantage of the rectal temperature measurement, not the bladder meas- urement. And finally, we had a small number of measurements in patients who were severely hypothermic or hyper- thermic. Hence, further studies with larger sample sizes including hypo and hyperthermic patients might further strengthen the results of this study. 5. CONCLUSIONS In conclusion, this study demonstrates that for post- operative cardiac surgery patients, bladder temperature measurement performs equal or better than rectal tem- perature measurement. Diuresis did not interfere with bladder temperature measurement. We believe that both from the patient’s perspective and from the nursing perspective, bladder temperature measurement has sufficient advantages in terms of ac- curacy, patient comfort and cost to be used in place of the cumbersome rectal temperature measurements in these patients. REFERENCES [1] Robb, P.J. and Shahab, R. (2001) Infrared transtympanic temperature measurement and otitis media with effusion. Internal Journal Pediatric Otorhinolaryngology, 59, 195- 200. doi:10.1016/S0165-5876(01)00482-7 [2] Mravinac, C.M., Dracup, K. and Clochesy, J.M. (1989) Urinary bladder and rectal temperature monitoring during clinical hypothermia. Nursing Research, 38, 73-76. doi:10.1097/00006199-198903000-00005 [3] Pompei, F. and Pompei, M. (1996) Physicians’ reference Copyright © 2012 SciRes. OPEN ACCESS ![]() H. Wollerich et al. / Open Journal of Nursing 2 (2012) 307-310 Copyright © 2012 SciRes. 310 OPEN ACCESS handbook on temperature. Exergen Corporation, Boston. [4] Bartlett, E.M. (1996) Temperature measurement: Why and how in intensive care. Intensive Critical Care Nurse, 12, 50-54. doi:10.1016/S0964-3397(96)81698-3 [5] Horrow, J.C. and Rosenberg, H. (1988) Does urinary catheter temperature reflect core temperature during car- diac surgery? Anesthesiology, 69, 986-989. doi:10.1097/00000542-198812000-00037 [6] Lefrant, J.Y., Muller, L., de La Coussaye, J.E., Benbabaali, M., Lebris, C., Zeitoun, N., et al. (2002) Temperature measurement in intensive care patients: Comparison of urinary bladder, oesophageal, rectal, axillary, and ingui- nal methods versus pulmonary artery core method. Inten- sive Care Medicine, 29, 414-418. doi:10.1007/s00134-002-1364-9 [7] Erickson, R.S. and Meyer L.T. (1994) Accuracy of infra- red ear thermometry and other temperature methods in adults. American Journal of Critical Care, 3, 40-54. [8] Giuliano, K.K., Scott, S.S., Elliot, S. and Giuliano, A.J. (1999) Temperature measurement in critically ill orally intubated adults: A comparison of pulmonary artery core, tympanic, and oral methods. Critical Care Medicine, 27, 2188-2193. doi:10.1097/00003246-199910000-00020 [9] Sato, H., Yamakage, M., Okuyama, K., Imai, Y., Iwashita, H., Ishiyama, T. and Matsukawa, T. (2008) Urinary blad- der and oesophageal temperatures correlate better in pa- tients with high rather than low urinary flow rates during non-cardiac surgery. European Journal of Anaesthesiol- ogy, 25, 805-809. doi:10.1017/S0265021508004602 [10] Langham, G.E., Maheshwari, A., Contrera, K., You, J., Mascha, E. and Sessler, D.I. (2009) Noninvasive tem- perature monitoring in postanesthesia care units. Anesthe- siology, 111, 90-96. doi:10.1097/ALN.0b013e3181a864ca [11] Lilly, J.K., Boland, J.P. and Zekan, S. (1980) Urinary bladder temperature monitoring: A new index of body core temperature. Critical Care Medicine, 8, 742-744. doi:10.1097/00003246-198012000-00010 [12] Robinson, J., Charlton, J., Seal, R., Spady, D. and Joffres, M.R. (1998) Oesophageal, rectal, axillary, tympanic and pulmonary artery temperatures during cardiac surgery. Ca- nadian Journal Anaesthesia, 45, 317-323. doi:10.1016/S0022-3476(98)70067-8 [13] Fallis, W.M. (2002) Monitoring urinary bladder tem- perature in the intensive care unit: State of the science. American Journal of Critical Care, 11, 38-45. [14] Fulbrook, P. (1993) Core temperature measurement: A comparison of rectal, axillary and pulmonary artery blood temperature. Intensive Critical Care Nurse, 9, 217-225. doi:10.1016/S0964-3397(05)80002-3 [15] Earp, J.K. and Finlayson, D.C. (1991) Relationship be- tween urinary bladder and pulmonary artery temperatures: A preliminary study. Heart Lung, 20, 265-270. [16] Earp, J.K. and Finlayson, D.C. (1992) Urinary bladder/ pulmonary artery temperature ratio of less than 1 and shi- vering in cardiac surgical patients. American Journal of Critical Care, 1, 43-52. [17] Moorthy S.S., Winn, B.A., Jallard, M.S., Edwards, K. and Smith, N.D. (1985) Monitoring urinary bladder tempera- ture. Heart Lung, 14, 90-93. ABBREVIATIONS CABG: Coronary artery bypass grafting; ICU: Intensive care unit; Tb: Bladder temperature; Tpa: Pulmonary artery temperature; Tr: Rectal temperature; PAC: Pulmonary artery catheter; SD: Standard deviation. |